Starting and Stopping Measurement and Recording
In situations where the MX100 cannot be connected to a PC, this function can be used to
acquire data on the MX100 by itself.
You can start and stop measurement and saving of data to the CF card by holding down
the CF card control switch for five seconds (the 5s-SW function).
Changing Modes with 5s-SW
CF card not inserted
CF card inserted
5s-SW
5s-SW
Idle Measrue
AlarmDO
5s-SW
Immediately
(after 1 s)
5s-SW
Idle Measrue
AlarmDO
Backup
CF
AlarmDO
If measurement fails to start, [ ] appears on the 7-segment LED for one second.
CAUTION
Do not use the 5 s-SW function when the MX100 and PC software are connected.
The PC software will experience an error.
Note
You can connect to the PC software after starting saving of data to the CF card using the
5 s-SW function.
The CF card can only be removed or inserted while in idle mode.
Operation When the CF Card Is Not Inserted
Normal Operation
Mode
Date
AlarmDO
Power On
5s-SW
Measure
Update Update
Idle
5s-SW
5s-SW
(Hold)
Hold
Operation When Power Cut and PC Software Connected
Connect to PC software
Save on PC
Record
(Start)
Auto restore
AlarmDO
PC File
Stop Start
Power On 5s-SW
Data
AlarmDO
Mode
Power Off On
MeasureMeasure Measure
UpdateUpdate Update
Hold
Data
Idle
5.2 Explanation of Functions