User Manual

4-30
IM MX100-01E
External Dimensions
Unit: mm
57
131
150.8
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.
Setting Value of Span and Preset, and Specified Value of User Output
PWM output setting value
span and preset value
Specified Value of User Output
(count)
0.000% 0
20.000% 20000
40.000% 40000
60.000% 60000
80.000% 80000
100.000% 100000
Output Span Setting
Span setting data when setting transmission output
Target output [%] 0.000 50.000 100.000
PC data* 0 50000 100000
* PC data refers to internal data between the PC and the MX100 main unit.
Handling Abnormal Data
Types of Abnormal Data Output Values
Abnormal data upon startup Preset value or previously held value can be selected
Data upon error occurrence Preset value or previously held value can be selected
When input channels differ Preset value or previously held value can be selected
+OVER Duty 100.000% output
–OVER Duty 0% output
±OVER determination
For command output, ±OVER determination is performed on the PC, and if the limit
value is exceeded, it is clipped at that limit value. For retransmission output, OVER is
determined by the MX100 main unit.
±OVER conditions
When duty exceeds 0.000-100.000%
When the input channel is ±OVER in the case of transmission output
4.11 8-CH, Medium-Speed PWM Output Module (MX120-PWM-M08) Specifications