2.10.7 Power ON Output Value Setting
Wire the units in the same manner as in the “Pulse Interval Accuracy Test,” enter all
settings (PWM output settings), then confirm that the criteria are met.
The wiring diagram for the pulse interval accuracy test shows an example of wiring CH5.
Actual measurment should be performed on all channels.
Use an external power supply of 5 V, and load resistance of 100 Ω.
PWM Outupt Settings
<MX100 Settings>
Connect to the PC, then enter the following settings using the MX100 standard software.
When using the MX100/DARWIN API for settings, enter settings that are equivalent to
the following.
Item Setting Remarks
Enable/disable Enabled
Select the source action User output
Span Need not specify
Output upon power ON Preset value Sets the output at power ON as a preset value.
Output upon error Preset value Sets the output upon error as a preset value.
Preset value 50.000 % Set 50% duty.
Resolution 1 ms
Pulse interval 200 ms
Other settings Not required
<MW100 Settings>
Using the MW100 user’s manual or communication commands, power cycle the MW100
main unit or equivalent, then measure the output on all channels.
Item Setting Remarks
Output type PWM
Output method COM Command output
Output range 1 ms
Span Need not specify
Pulse interval 200
Preset value 50.000 % Set 50% duty.
Reference channel Not required
Output Operation Setting
Item Setting
Power ON PRESET
Error occurence PRESET
Criteria (for the MX100 and MW100)
Power cycle the MX100 main unit or equivalent, then measure the output values without
entering any settings using the PC.
Use a trigger level of 2.5 V. In this case, confirm that the output voltages of all channels
are within the allowable ranges.
Pulse Interval Pulse Duty
200 ms ±16S Duty 50% ±0.003%