Manual

2-46
SM MX100-01E
<MW100 Settings>
Referring to the MW100 user’s manual for browser operations, or by using
communication commands, power cycle the MW100 main unit or equivalent, then
measure the output on all channels.
Output Range Settings
Item Setting Remarks
Output type PWM
Output method COM Command output
Output range 1 ms
Span Need not specify
Pulse interval 10
Preset value 0.000% Duty 0%
Reference channel Not required
Output Operation Setting
Item Setting
Power ON PRESET
Error occurence PRESET
Criteria (for the MX100 and MW100)
PWM Output Setting Criteria
100.0 % 3.8 V or more
2.10.5 Pulse Interval Accuracy Test
Wire the units as in the figure below, enter all settings (PWM output settings), then
confirm that the criteria are met.
The figure below shows an example of wiring CH5. Actual measurment should be
performed on all channels.
Use an external power supply of 5 V, load resistance of 100 Ω, and a counter trigger level
of 2.5 V.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
+
-
100Ω
Counter
+
-
5V
V+
-
5Ch
V+
-
6Ch
V+
-
7Ch
V+
-
8Ch
+
-
Vext
+
-
Vext
V+
-
4Ch
V+
-
3Ch
V+
-
2Ch
V+
-
1Ch
Puls interval check and puls duty check wiring diagram
DC
Power
supply
2.10 8-CH, Medium-Speed PWM Output Module