Testing
1
2
3
4
5
6
7
2.6.5 Measurement Accuracy Test
Procedure
1.
Turn the power ON again, then allow the instrument to warm up for at least 10 minutes.
2.
Input DC voltage and resistance to channel 2 of the MX110-V4R-M06, then
confirm that the specifications are met at each range.
• Settings on Module under Test: Set the range and measurement interval
according to the table.
Measurement Interval Duration
100 ms 5 seconds or more
500 ms 10 seconds or more
• Test the module’s input voltage while measuring the DC voltage generator’s
output on a digital multimeter.
• If calibration was performed, power cycle the instrument before the test.
Confirm that the measured results are within the judgement criteria range in the
chart below.
Range Input Voltage Measurement Interval Specification
20 mV +20.000 mV 500 msec ±(1% of rdg + 25 digits)
0.000 mV
-20.000 mV
60 mV +60.00 mV 500 msec ±(0.1% of rdg + 10 digits)
0.00 mV
-60.00 mV
200 mV +200.00 mV 500 msec
0.00 mV
-200.00 mV
1 V +1.0000 V 500 msec
0.0000 V
-1.0000 V
2 V +2.0000 V 500 msec
0.0000 V
-2.0000 V
6 V +6.000 V 500 msec
0.000 V
-6.000 V
20 V +20.000 V 500 msec
0.000 V
-20.000 V
100 V +100.00 V 500 msec
0.00 V
-100.00 V
Cu10 GE
High Resolution
+1.326 Ω/-200.0°C 500 msec ±(0.2% of rdg + 5°C)
+9.036 Ω/0.0°C
+20.601 Ω/+300.0°C
Pt100 +18.52 Ω/-200.0°C 500 msec ±(0.1% of rdg + 1.5°C)
+100.00 Ω/0.0°C
+313.71 Ω/+600.0°C
Pt1000 +185.2 Ω/-200.0°C 500 msec ±(0.1% of rdg + 1.5 digits)
+1000.0 Ω/0.0°C
+3137.1 Ω/+600.0°C
20 Ω +20.000 Ω 500 msec ±(0.1% of rdg + 25 digits)
0.000 Ω
200 Ω +200.00 Ω 500 msec ±(0.1% of rdg + 15 digits)
+10.00 Ω
0.00 Ω
2 kΩ +2000.0 Ω 500 msec ±(0.1% of rdg + 10 digits)
0.0 Ω
2 V +2.0000 V 100 msec ±(0.05% of rdg + 5 digits)
0.0000 V
-2.0000 V
2.6 Tests of the 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06)