Manual

2-14
SM MX100-01E
2.5 Tests of the 10-CH, Medium-Speed Universal
Input Module (MX110-UNV-H10)
Tests
2.5.1 Insulation Resistance Test
2.5.2 Withstanding Voltage Test
2.5.3 Excessive Input Test
2.5.4 ID Check
2.5.5 Measurement Accuracy Test
2.5.6 Between-Channel Error Test
2.5.7 Burnout Test
2.5.8 Reference Junction Compensation Accuracy Test
Instruments Used
Instrument Standards, Recommended Products Etc.
MX100, MW100, or equivalent A device that can connect with and perform communications with
the MX110 10-ch Medium-Speed Universal Input Module (Ex.
MX100 + MX150 or MW100 + MX150)
MX150 Base Plate
Front TR Assembly (When with
the /NC option)
Product number: B8720BE
A device that can receive input independently on each channel
Screw Term Assembly & Cable Product number: B8720BH
Used to confirm the ID setting. If the ID is correct, a device with an
altered connector or the like can be substituted.
PC
A device that can perform communication with the MX100/M\W100
or equivalent instrument
DC voltage generator FLUKE 5520A* or equivalent
Accuracy: 0.005% of setting + 1mV
A device capable of 100 V output or higher
Rheostat ALPHA ADR-1202
or equivalent
Accuracy: 0.005% of setting + 2 mΩ
A device with resolution of 1 mΩ or less
DMM Agilent 3458A or equivalent
Voltage: 0.005% of rdg + 1mV
Resistance: 0.01% of rdg + 1 mΩ
Insulation resistance tester DC500 V
Withstanding voltage tester AC 1 kV–4 kV
Temperature tank standardized
at 0°C
Coper Electronics ZC-114 or equivalent
0±0.01°C
* If a DC voltage generator meeting the standards above is not available, you can ensure that
the stated voltage is output by substituting a FLUKE 9100 or equivalent and monitoring the
output voltage using a DMM. Also, due to the output impedance, some voltage generators
cannot output all stated voltages, and at 0 V, there are many cases where the above
standard is not met, so 0 V can be obtained by shorting the terminals. It is recommended to
monitor when set to DOWN.
Measure the setting value and actual output value, then confirm that the above standards
are met.
0.2 Ω or less cannot be set on the ADR-1202, so 0 Ω can be achieved by shorting the
terminals.
Ensure that the module settings and measured values can be read in.