Notes on Measurement
Shunt resistance for current measurement is built in. Measurements are taken as DC
voltage values in the DC voltage range.
When displaying current values, convert voltage values to current values by using scaling
or other means.
Current value [A] = voltage value [V] / shunt resistance value [Ω]
Main Specifications
Style number: S1
No. of connectable points: 10
Type of measurement: DC current
Measurable range: 772081: −100 to 100 mA
772082: −30 to 30 mA
772083: −20 to 20 mA
Note that this depends on the DC voltage range used.
Measurement accuracy: Add the following values to the measurement accuracy of the
DC voltage range being used.
±0.15 % of rdg.
When scaling, add an additional 2 digits of computation error.
Connection type: Insulation between channels
Note that the NC terminal is common between channels.
Normal-mode voltage: Depends on the voltage range used.
Terminal type: Clamp
Applicable wire size: 0.14 to 1.5 mm2 (AWG26 to 16)
Effects of ambient temperature:
When the integral time is 16.67 ms or more, the effect per
changeof 10 °C in ambient temperature is as follows.
Within ±(0.075% of rdg. + 0.05% of range*)
* Indicates the DC voltage used.
Other: Other basic specifications conform with those of the 10-CH,
Medium-Speed Universal Input Module (MX110-UNV-M10).