Attaching the Terminal Plate
With the /NC option, begin from step 3.
1. Loosen the terminal cover attachment screw for the 10-CH Medium-Speed Universal
Input Module, then rmly ip over the terminal cover in the direction of the arrow in the
gure below and remove it.
2. Loosen the attachment screw of plate, then remove the terminal plate.
3. Attach the plate with clamp terminals for current, then fasten with the screw.
4. Attach the terminal cover that came with the plate with clamp terminals for current.
Terminal cover (standard accessory)
Plate with clamp terminals
(standard accessory)
Terminal cover attachment screw
Attachment screw
(loosed before pulling off the terminal plate)
1
2
1
2
5
4
8
7
11
10
14
13
1
2
5
4
8
7
11
10
14
13
2
5
4
8
7
11
10
14
13
1
2
5
4
8
7
11
10
14
13
Plate with clamp terminals
for current (fasten with screw)
Terminal cover that
comes with the plate with
screw terminals for current
3
4
Notes on Wiring and Calibration
It is recognized by the PC software as a universal input plate with clamp terminals.
Also, the 10-CH Universal Input Module cannot be calibrated when the plate with clamp
terminals for current is attached. Calibrate the 10-CH Universal Input Module with the
terminal plate attached that was attached during shipment, or with accessory terminal
772061, 772063, or 772080 (sold separately) attached.