User guide

4
IM MX100-74E
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsofthe10-CH,High-SpeedDigitalInputmoduleandtheoutput
terminalsofthe10-CH,Medium-SpeedDigitalOutputmodule.Doingsocan
damagethemodules.
• Maximuminputvoltage
 10-CH,High-SpeedDigitalInputModule:±10VDC(-D05),±50VDC(-D24)
 10-CH,Medium-SpeedDigitalOutputmodule:±250VACor250VDC
• Maximumcommonmodevoltage
 Input/outputterminaltoground:250VACrms(50/60Hz)
• Donotapplyavoltageexceedingthevalueindicatedbelowtotheinput
terminalsoftheanalogoutputmodulesorthePWMoutputmodules.Doingso
candamagethemodules.
• Maximumcommonmodevoltage
 Outputterminaltoground:250VACrms(50/60Hz)
• TheMX100andMW100areanovervoltagecategoryII(IEC61010-1)
instrument.MeasurementcategoryII(IEC61010-2-30)isappliedtotheuniversal
inputmodules,theDCV/TC/DIinputmodule,theFour-wireRTDresistanceinput
module,andthestraininputmodule.
Wiring Procedure, Specifications, and Other Items Related to the Input/Output Modules
SeetheMX100/MW100DataAcquisitionUnitInstallationandConnectionGuide(IM
MX100-72E)providedwiththemainmodule,ortheMX100DataAcquisitionUnitUsers
Manual(IMMX100-01E)ortheMW100DataAcquisitionUnitUsersManual(IM
MW100-01E)containedinthemanualCD-ROM.