Timer (<timerNo>)
Up to eight timers can be specified. The timer is specified using a number between 1
and 8.
Select the timer operation from the choices below. Select Off to disable the timer.
Edge: Returns 1.0 when performing computation after the ON time elapses after the
beginning of each interval. Specify the interval for detecting the condition and
the ON time in ms.
Level:
Returns 1.0 from the time after the ON time elapses un
til after the OFF time
elapses. Specify the interval for detecting the condition, ON time, and OFF
time.
The selectable range of the interval for detecting the condition is 1 s to 86400000 s =
1000 days. The ON/OFF time is specified using the elapsed time from the beginning
of the interval.
Timer Level Operation
• When ON < OFF
• When ON > OFF
ON setting
OFF setting
t
Interval
Returns 1.0
Returns 1.0
Interval
ON OFF ON OFF
ON setting
OFF setting
t
Interval
Returns 1.0Returns 1.0 Returns 1.0
Interval
OFF ON OFF ON
2.5 Setting Computations (Setting the Computation Channels)