Owner's manual

2-24
IM MX180-01E
On the channels of the input module, select from the modes below according to the input
signal. For details of the measurable range, measurement accuracy, and resolution, see
chapter 4, “Specification” in the MX100 Data Acquisition Unit Users Manual (IM MX100-
01E).
VOLT (DC voltage)
Select the measurement range from 20 mV, 60 mV, 200 mV, 2 V (default), 6 V, 20 V
,
or 100 V.
Note
Current input is possible by attaching a shunt resistor to the input terminal and converting to voltage
input. The shunt resistors in the table below are available for purchase separately. For example, a
250 Ω shunt resistor is used to convert the signal in the range of 1 to 5 V for 4 to 20 mA input.
Part Name Model Resistance
Shunt resistor
(for clamp terminal)
415920
415921
415922
250 Ω ± 0.1%
100 Ω ± 0.1%
10 Ω ± 0.1%
Shunt resistor
(for screw terminal)
438920
438921
438922
250 Ω ± 0.1%
100 Ω ± 0.1%
10 Ω ± 0.1%
TC (thermocouple), on the Universal Input Module and 30-CH, Medium-Speed
DCV/TC/DI Input Module
Select
the thermocouple type (referred to as Range in the setup) from Type-R (default),
Type-S, Type-B, Type-K, Type-E, Type-J, Type-T, Type-N, Type-W, Type-L, Type-U, or
KpvsAu7Fe.
RTD1 (resistance temperature detector, measured current: 1 mA), on the 4-CH,
High-Speed Un
iversal Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100 (high resolution Pt100), HQ JPt100 (high-resolution JPt100), Ni100:SAMA,
Ni100:DIN, or Ni120.
RTD1 (resistance temperature detector, measured current of 1 mA), on the
10-CH, Medium-Speed Un
iversal Input Module/6-CH, Medium-Speed Four-Wire
RTD Resistance Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100 (high-resolution Pt100), HQ JPt100 (high-resolution JPt100), Ni100:SAMA,
Ni100:DIN, Ni120, Pt50, Cu10:GE, Cu10:L&N, Cu10:WEED, Cu10:BAILEY, or J263B.
RTD2 (resistance temperature detector, measured current: 2 mA), on the 4-CH,
High-Speed Universal Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100 (high-resolution Pt100), HQ JPt100 (high-resolution JPt100), Pt50, Cu10:
GE, Cu10:L&N, Cu10:WEED, Cu10:BAILEY, or J263B.
The RTD2 selection is not available on the 10-CH, Medium-Speed Universal Input
Module and the 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module.
RTDEX (resistance temperature detector
, measured current: 0.25 mA), on the
6-CH, Medi
um-Speed Four-Wire RTD Resistance Input Module
Select the RTD type (referred to as Range in the setup) from Pt500 (default setting) or
Pt1000.
OHM (resistance), on the 6-CH, Medium-Speed Four-Wire RTD Res
istance Input
Module
Select a measurement range of 20 ohm (measured current: 1 mA), 200 ohm (measured
current: 1 mA), or 2 kohm (measured current: 0.25 mA).
2.4 Setting the Measurement Conditions (Setting the Measurement Channels)