5.2 Error Messages and Their Corrective Actions
Error Messages on the Integration Monitor
Error Code Message Corrective Action/Description
Reference
Page
IP address is invalid. Enter the correct IP address. 2-7
Subnet mask is invalid. Enter the correct subnet mask. 2-7
Default gateway is invalid. Enter the correct gateway address. 2-7
Failed to update. Attempted to change the IP address while the connection is
established, or the MX100 you wish to configure is connected
by another software program. Disconnect the MX100.
-
Search MX100s nearby again. Click the Search button again on the Connection screen, and
change the settings.
2-1, 2-5
Broken modules exist. Remove the input/output modules one by one and determine
the broken module, then replace it.
*
Some modules are not calibrated
correctly.
Recalibrate. 4-4
Undetermined directory. The save destination of the record file cannot be found. Check
whether the save destination exists.
2-16
Write error The record file cannot be created. Check whether free space
on the destination drive exists, whether write permissions have
been granted, whether the disk is damaged, and whetherthe
file system is corrupt.
-
E5102 Failed to format CompactFlash. Check whether communication with the MX100 is being
performed.
Insert a CF card correctly.
2-7
*
E5103 Invalid serial number. Please
reinstall the software.
Enter the correct serial number. 1-5
E5104 Failed to create folder. The directory of the folder to create is not correct. Check the
file save destination.
2-16
E5105 Invalid folder name. Please enter
a new name.
Entered folder name is incorrect. Enter a correct folder name. -
E5106 Cannot contain any of the
following characters:
\\ / : , ; * ? " < > | CON PRN AUX
CLOCK$ NUL COM1-9 LPT1-9
Entered folder name includes illegal characters. Enter a correct
folder name.
-
E5107 Compact Flash slot is empty. Insert a CF card correctly. *
W5131 Update the display settings. The display setup is initialized according to the measurement
and computation channel settings.
2-43
W5132 Writing data files Currently. During finalization, do not close MX100 Standard. 2-51
W5133 Cancel the current settings and
create a new project?
Settings are discarded and the MX100 Standard is restored to
its default state.
2-6
W5134 The current channel setup
contains invalid use of reference
channels.
Click No and change the settings of the improper channels. To
disable the channels with improper settings, click Yes.
2-25 to 2-27
W5135 In order to balance the strain
channels, it's necessary for
channels to be set up correctly.
The current channel setup
contains invalid use of reference
channels.
Click No and change the settings of improper channels. To
disable the channels with improper settings, click Yes.
2-25 to 2-27
W5136 Format Compact Flash? The CF card is formatted. 2-11
W5137 Stop Record? Data saving stops. 2-51
W5138 Module settings which do not
match the current software
configuration will be initialized.
If you click Yes, settings of modules that differ between the
MX100 module configuration and the configuration recognized
by MX100 Standard will be initialized. To not reconfigure, click
No.
2-10
W5139 Some detected modules do
not match the current software
configuration. Rebuild based on
the current MX hardware setup?
Click Yes reconfigure. To not reconfigure, click No. 2-9
* See the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).