Owner's manual
4-11
IM MX180-01E
Calibrator
3
2
1
4
5
App
Index
4.2 Calibration Procedure
• 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06)
Measurement Range to Be Calibrated Input Value 1 Input Value 2
20 mV 0 mV on CH1 20 mV on CH2
60 mV 0 mV on CH1 60 mV on CH2
200 mV 0 mV on CH1 200 mV on CH2
1 V 0 V on CH1 1 V on CH2
2 V 0 V on CH1 2 V on CH2
6 V 0 V on CH1 6 V on CH2
20 V 0 V on CH1 20 V on CH2
100 V 0 V on CH1 100 V on CH2
RTD(1 mA)20 mV 0 Ω on CH3 20 Ω on CH4
RTD(1 mA)60 mV 0 Ω on CH3 60 Ω on CH5
RTD(1 mA)200 mV 0 Ω on CH3 200 Ω on CH6
RTD(1 mA)600 mV 0 Ω on CH3 300 Ω on CH4
RTD(0.25 mA)600 mV 0 Ω on CH3 2400 Ω on CH5
RTD(0.25 mA)1 V 0 Ω on CH3 3000 Ω on CH6
• 4-CH, Medium-Speed Strain Input Module (MX112-B12-M04, MX112-B35-M04,
MX112-NDI-M04)
Measurement Range to
Be Calibrated
Connection 1
(Zero Calibration)
Connection 2
(Full Calibration)
2000 µSTR (strain) 120.000 Ω on R1 to R4 of CH2
120.000 Ω on R1 to R3 of CH2
and 117.154 Ω on R4
20000 µSTR (strain) 120.000 Ω on R1 to R4 of CH2
120.000 Ω on R1 to R3 of CH2
and 113.010 Ω on R4
200000 µSTR (strain) 120.000 Ω on R1 to R4 of CH2
120.000 Ω on R1 to R3 of CH2
and 80,000 Ω on R4
Use a 4-gauge method connection when performing calibration. For information on
this connection, see section 2.4 of the MX100 Data Acquisition Unit User’s Manual (IM
MX100-01E).
• 8-CH, Medium-Speed
Analog Output Module (MX120-VAO-M08)
Adjust so that all channels output 0 V and 10 V. This differs from other input modules.
Range to Be Calibrated Output Value 1 Output Value 2
10 V 0 V 10 V
Note
• For a description of the measurement range and accuracy of each input module, see
chapter 4 in the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).
• For a description of the calibration procedure and wiring, see section 3.3 of the MX100 Data
Acquisition Unit User’s Manual (IM MX100-01E).