Owner's manual

4-10
IM MX180-01E
4.2 Calibration Procedure
Explanation
Calibrated Parameters
4-CH, High-Speed Universal Input Module (MX110-UNV-H04)
Calibration is carried out on a channel-by-channel basis.
Measurement Range to Be Calibrated Input Value 1 Input Value 2
20 mV 0 mV 20 mV
60 mV 0 mV 60 mV
200 mV 0 mV 200 mV
1 V 0 V 1 V
2 V 0 V 2 V
6 V 0 V 6 V
20 V 0 V 20 V
100 V 0 V 100 V
RTD(1 mA)200 mV 0 Ω 200 Ω
RTD(1 mA)600 mV 0 Ω 300 Ω
RTD(1 mA)1 V 0 Ω 500 Ω
RTD(2 mA)60 mV 0 Ω 30 Ω
RTD(2 mA)200 mV 0 Ω 100 Ω
RTD(2 mA)600 mV 0 Ω 300 Ω
RTD(2 mA)1 V 0 Ω 250 Ω
10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10)
Measurement Range to Be Calibrated Input Value 1 Input Value 2
20 mV 0 mV on CH1 20 mV on CH2
60 mV 0 mV on CH1 60 mV on CH2
200 mV 0 mV on CH1 200 mV on CH2
1 V 0 V on CH1 1 V on CH2
2 V 0 V on CH1 2 V on CH2
6 V 0 V on CH1 6 V on CH2
20 V 0 V on CH1 20 V on CH2
100 V 0 V on CH1 100 V on CH2
RTD(1 mA)60 mV 0 Ω on CH3 60 Ω on CH6
RTD(1 mA)200 mV 0 Ω on CH3 200 Ω on CH4
RTD(1 mA)600 mV 0 Ω on CH3 300 Ω on CH5
30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30)
Measurement Range to Be Calibrated Input Value 1 Input Value 2
20 mV 0 mV on CH1 20 mV on CH2
60 mV 0 mV on CH1 60 mV on CH2
200 mV 0 mV on CH1 200 mV on CH2
1 V 0 V on CH1 1 V on CH2
2 V 0 V on CH1 2 V on CH2
6 V 0 V on CH1 6 V on CH2
20 V 0 V on CH1 20 V on CH2
100 V 0 V on CH1 100 V on CH2