User’s Manual MX100 Standard Software IM MX180-01E 4th Edition
Thank you for purchasing the MX100 Data Acquisition Unit. This user’s manual contains useful information about the functions and operating procedures of the MX100 Standard Software and lists the handling precautions of the software. To ensure correct use, please read this manual thoroughly before beginning operation. After reading the manual, keep it in a convenient location for quick reference whenever a question arises during operation.
Revisions • 1st Edition: • 2nd Edition: • 3rd Edition: ii May, 2003 November, 2004 February, 2008 • 4th Edition: March 2012 IM MX180-01E
Terms and Conditions of the Software License NOTICE - PLEASE READ CAREFULLY BEFORE USE Thank you very much for purchasing this medium containing a software program and related documentation provided by Yokogawa Electric Corporation (hereinafter called “Yokogawa”), and the program contained, embedded, inserted or used in the medium (hereinafter called the “Yokogawa Software Program”).
Terms and Conditions of the Software License 4.3 When Yokogawa decides in its own judgement that it is necessary, Yokogawa may from time to time provide the Licensee with Revision upgrades and Version upgrades separately specified by Yokogawa (hereinafter called “Updates”). 4.4 Notwithstanding the preceding Paragraph 4.3, in no event shall Yokogawa provide Updates where the Licensee or any third party conducted renovation or improvement of the Licensed Software. 4.
1 Contents Terms and Conditions of the Software License................................................................................ iii Chapter 1 Chapter 2 Before Using the Software 1.1 1.2 1.3 1.4 Functional Explanation of the MX100 Standard Software................................................. 1-1 System Requirements........................................................................................................ 1-4 Setup Procedures of the MX100 Standard Software.......................
Chapter 1 Before Using the Software 1.1 Functional Overview The software can be used when the PC and the MX100 are connected in a one-to-one configuration. To connect multiple MX100s to a PC, you need to use MXLOGGER, a software program sold separately.
1.1 Functional Explanation of the MX100 Standard Software • Viewer Enables you to display measured and computed data that has been saved, read values and perform statistical computation over an area using cursors, and convert the measured and computed data into various file formats such as Excel. • Calibrator Enables you to calibrate the MX100 input/output modules. Integration Monitor The main operations are carried out using the six windows shown below.
1.1 Functional Explanation of the MX100 Standard Software • Channel setup screen • Display setup screen For each display group, you can set the channel to be displayed, display scale, display zone, trip point (the value of interest in the waveform display), display color, and other parameters.
1.2 System Requirements OS (Operating System) Run the software under any of the following operating systems.
1.3 1 Before Using the Software Setup Procedures of the MX100 Standard Software 2 The following procedures are for installing the software on Windows XP. 1. Start Windows. Log onto Windows as an administrator (Otherwise, you will not be able to install the software). This is the same when using an OS other than Windows XP. 3 When Auto Run Is Enabled 2. Insert the CD-ROM in the CD-ROM drive. 4 The setup program automatically starts. Proceed to step 4. When Auto Run Is Disabled 5 2.
1.3 Setup Procedures of the MX100 Standard Software 7. Select the setup type to install, and click Next. The default destination to install is [C:\MX100 Standard]. If you click Custom check box, you can select the installation location and installed functions. Note that if you select MX100 Standard in the installation function, MX100 Viewer will also be installed. Follow the onscreen instructions. 8. Click Install to start the installation. 9. Click Finish to finish the installation.
1.3 Setup Procedures of the MX100 Standard Software 1 Explanation When the software program is installed properly, the folder MX100 Standard Software is created in the specified directory (C:\ by default). MX100 Standard is registered in the program list and MX100 Standard (referred to as the Integration Monitor in this operation guide), MX100 Viewer, and MX100 Calibrator are registered under it.
1.4 Starting and Exiting Software Programs Procedure Starting the Software Programs As shown in the figure below, click the Start button, point to Programs, point to MX100 Standard, and choose MX100 Standard (referred to as the Integration Monitor in this manual), MX100 Viewer, or MX100 Calibration. Exiting the Software From the File menu, choose Exit.
Chapter 2 2.1 Integration Monitor 1 Connecting and Disconnecting the MX100 2 Procedure Note Connecting the MX100 and the PC to a mission-critical network before an appropriate IP address is assigned to the MX100 may adversely affect other connected devices on the network. Connecting to an MX100 That Has Not Been Configured for the Network Integration Monitor Before carrying out the procedure below, turn ON the power to the relevant MX100 and connect the MX100.
2.1 Connecting and Disconnecting the MX100 3. Click the MX100 icon that appeared. In the box containing the Edit Network setup button, the icon and the initial network information of the selected MX100 appear. Even if you click the Connect button in this condition (with the initial IP address), connection cannot be made. Default network information Clicking here causes the 7-segment LED of the corresponding MX100 to display --CALL-Click the MX100 icon 4. Click the Edit Network setup button.
2.1 Connecting and Disconnecting the MX100 1 7. Click the MX100 icon in the box containing the Search button. In the box containing the Edit Network setup button, the icon and the network information of the configured MX100 appear. 2 Integration Monitor 3 4 Click here 5 Displayed with the changes applied App Index 8. Click the Connect button in the box containing the Edit Network setup button. If the MX100 is connected, the Monitor screen appears and measurement begins.
2.1 Connecting and Disconnecting the MX100 Connecting to an MX100 That Has Been Configured for the Network 1. Start the Integration Monitor. If the MX100 that was connected when the program was started the previous time is still connected, a connection is automatically established with that MX100, the Monitor screen opens, and measurement starts. To connect to another MX100, click Connection located at the top section of the Monitor screen to show the Connection screen and click Disconnect. 2.
2.1 Connecting and Disconnecting the MX100 1 Changing the Network Configuration 1. With the MX100 icon and network information displayed in the box containing the Edit Network setup button on the Connection screen, click the Edit Network setup button. 2. Refer to the explanations in this section and enter the required information in the boxes. The Apply and Cancel buttons appear. 3 4 3. Click Apply. Click Cancel to return to the original settings.
2.1 Connecting and Disconnecting the MX100 Disconnecting the MX100 1. Click Connection to show the Connection screen. 2. Click the Disconnect button to the right of the IP Address or Host Name text box. When the connection is dropped, a message “Currently no MXs are connected” appears at the top section of the screen. To reconnect, click Connect.
2.1 Connecting and Disconnecting the MX100 1 Explanation Machine Name To distinguish individual MX100s, you can enter a name that is easy to identify using up to sixty-four characters (not required). Enter a new IP address to assign to the MX100. The default address is 127.1.1.XX (where XX is a unique number for the device). This address cannot be used even when the PC and the MX100 are connected in a one-to-one configuration. You must change to a different address such as 192.168.1.
2.1 Connecting and Disconnecting the MX100 Model Displays MX100. Option Displays /DS or None. Version Firmware version number of the main module. The version number shows the style number (before the decimal point) and revision number (after the decimal point). Serial ID The instrument number (NO.) marked on the name plate of the main module. MAC Address Displays the MAC address that is assigned when the MX100 is shipped from the factory (unique ID number for identifying the Ethernet card).
2.2 Configuring the System, Setting the Write Mode of the CF Card, and Formatting the CF Card Procedure If you change the attachment positions or type of the input/output modules you must carry out the procedure below. 3 Note • • If you change the MX100 input/output modules, a message, “Some detected modules do not match the current software configuration. Rebuild based on the current MX hardware setup?” appears when you start the Integration Monitor.
2.2 Configuring the System, Setting the Write Mode of the CF Card, and Formatting the CF Card 3. Click the Current Configuration tab followed by the Reconfigure based on actual config. button. A dialog box with the message “Module settings which do not match the current software configuration will be initialized.” appears.
2.2 Configuring the System, Setting the Write Mode of the CF Card, and Formatting the CF Card 1 MX100 Response Check 1. Click System to show the System setup screen. 2 2. Click the Execute button to the right of Response Check. The 7-segment LED on the main module of the connected MX100 displays “--CALL--” Integration Monitor 3 4 5 App Click here Index Setting the Write Mode and Formatting the CF Card 1. Click System to show the System setup screen. 2.
2.2 Configuring the System, Setting the Write Mode of the CF Card, and Formatting the CF Card While the CF card is being formatted, the 7-segment LED of the main module displays the following. Explanation Items That Can Be Confirmed on the System Screen • Unit Setup You can check the model, version, serial ID, option, temperature unit, and the A/D integration time.
2.
2.3 Setting the Acquisition Conditions of the Measured/Computed Data Procedure Setting the Monitor Interval and Record Interval of Measured Data 1. Click Acquisition to show the Acquisition setup screen. Click here 2. Drag and drop the input module icon within the Measurement Group 1, Measurement Group 2, or Measurement Group 3 frame. Drag the input module with the same monitor interval to the same measurement group.
2.3 Setting the Acquisition Conditions of the Measured/Computed Data 1 Setting the Record Interval of Computed Data 1. Click Acquisition to show the Acquisition setup screen. 2. Enter the record interval (integer multiplication factor) in the Rec. Interval text box. 2 Integration Monitor 3 4 5 App Set using an integer multiple of the monitor interval for each computation group (up to 128) The recording to the data file is carried out at (monitor interval × multiplication factor).
2.3 Setting the Acquisition Conditions of the Measured/Computed Data Setting the Record File 1. Click Acquisition to show the Acquisition setup screen. 2. Enter the save destination of the file in the Folder box. If you click the folder icon, the Browse for Folder dialog box opens, and you can select the save destination folder. To do so, click the New Folder button, enter the folder name to create a new folder within the save destination folder, and then set it as the save destination. 3.
2.3 Setting the Acquisition Conditions of the Measured/Computed Data 1 Explanation Monitor Interval Note The shortest interval that you can specify for the monitor interval is the longest of the shortest measurement intervals of all the input modules assigned to the same measurement group. For example, if an input module with the shortest measurement interval of 10 ms and an input module with 100 ms are assigned, 100 ms is the shortest interval that can be specified for the group.
2.3 Setting the Acquisition Conditions of the Measured/Computed Data Record Stop Condition Select the condition for stopping the recording from the following: • Continuous Immediately stops recording when the Stop button on the Monitor screen is clicked or Stop is chosen from the Acquisition menu. • Fixed Time Stops recording at the specified date/time. Set the date/time as “year/month/day/ hours:minutes:seconds” If the specified time has already been passed, the operation is the same as Manual.
2.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) Setting the Scale (Only When Scaling) 8. Select the Use check boxes under Scale to enable changing of the scale. 9. Select the decimal point position from the Point list box for Scale. 10. Click the Min box or Max box under Scale and enter the minimum or the maximum value of the scale. 11. Click the Unit box and enter the unit of the scaled value.
2.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) Set the type for alarms 1 to 4, the setting value, and the hysteresis (see section 2.9) Set the initial balance for strain input channels (only when the input mode is set to STR) Set the Initial Balance for Strain Input Channels (Only When the Input Mode Is Set to STR) 19. Under Strain function, click the Execute button under Initial balance or Reset.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) Note • You can execute initial balancing and reset the initial balance value on multiple recording channels at once by selecting the channels and clicking the collective execution button.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) On the channels of the input module, select from the modes below according to the input signal. For details of the measurable range, measurement accuracy, and resolution, see chapter 4, “Specification” in the MX100 Data Acquisition Unit User’s Manual (IM MX10001E). • VOLT (DC voltage) Select the measurement range from 20 mV, 60 mV, 200 mV, 2 V (default), 6 V, 20 V, or 100 V.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) • DI (digital input), on the 10-CH, Medium-Speed Digital Input Module Only the DI (digital input) setting is available for the input type. Select LEVEL (D10) (voltage input) for -D05, or CONTACT (D10) (non-voltage contact input) for the measurement range to match the input. For -D24, only LEVEL (24 V) is available.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) Setting the Reference Channel for Difference Input Set this item when making the difference between the measured value of the channel and the measured value of the reference channel the measured value (referred to as difference computation). The reference channel can be set to a channel that is scaled.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) Setting the Burnout Setting the RJC When the input mode is TC, select whether to use the internal reference junction compensation function of the input module or an external reference junction compensation function. (This setting is possible when the input mode is set to a mode other than DI, but the setting takes effect only when the mode is TC.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) The results of initial balancing are as follows. Status box display ----Successful Failed OVER Executing Status Not executed Initial balancing was executed successfully. Initial balancing failed. Initial balancing succeeded, but the output value exceeded the upper limit or fell below the lower limit, and the value was rounded to the upper limit value or the lower limit value.
2.4 Setting the Measurement Conditions (Setting the Measurement Channels) Generally, the range is as follows. Using the symbols that have appeared in the explanation thus far, after setting the scale maximum and minimum, the values are as follows. 3 When Listed at the Calibration Coefficient An example is given with a displacement gauge. • Rated input 20 mm • Calibration coefficient 0.
2.5 Setting Computations (Setting the Computation Channels) Procedure Click here Math Channel tab Minimize button Record check box Monitor check box Enter the equation (up to127 characters, not case sensitive) Equations that are appropriate are displayed in black Equations that are inappropriate are displayed in peach Specifies the decimal places for the minimum/maximum value of span Used for output channel scaling and the initial value of the span for the General Display screen. 1.
2.5 Setting Computations (Setting the Computation Channels) 1 2 Enter the tag number (up to 15 characters) 3 4 Enter the tag comment (up to 30 characters) Enter the label for identifying the constant Appropriate labels are displayed in black Inappropriate labels are displayed in peach Enter the constant value Alarm settings (see section 2.9) 5 App Index Note Sixty computation channels are always shown. Setting the Unit, Tag Number, and Tag Comment 8. Click the Unit, Tag No.
2.5 Setting Computations (Setting the Computation Channels) Importing Tag Numbers and Tag Comments You can import just the tag numbers and tag comments from the saved setup data and replace the current tag numbers and tag comments with the imported information. From the File menu, choose Import Tag. Then, select an MX100 project file (a file with the .mxp extension) in the Open dialog box to import the tag numbers and tag comments. Select the MX100 project file (file with .
2.5 Setting Computations (Setting the Computation Channels) 1 Setting Timers (Only When Using Timers in the Computations) 11. Click Action for a timer and select Edge or Level. When not using the timer, select Off (default). 2 12. Click the Interval box, and enter the interval. which the timer is to be turned ON. If Action was set to Level, click the On-period or Off-period box and enter the time over which the timer is turned ON or OFF. Integration Monitor 13.
2.5 Setting Computations (Setting the Computation Channels) Explanation Computation Overview You can enter computation equations using the constants, operators, and functions described below. The computed results can be displayed and recorded (saved) as computed data. Computation allows you to determine the average/maximum/minimum of a specified channel on a specified date/time, or output events (start/stop record, reset time, etc.) under specified conditions.
2.5 Setting Computations (Setting the Computation Channels) Manual Output, and User Output (DO Channels, AO Channels, and PWM Channels) Operators The operators listed below can be used.
2.5 Setting Computations (Setting the Computation Channels) Reference Functions and TLOG Functions The function below are used to retrieve measured values and alarm values. The functions are not case sensitive. For the format used to specify the , , and , see “Procedure” in this section.
2.5 Setting Computations (Setting the Computation Channels) 1 Arithmetic Functions Below are the arithmetic functions that are available. They are not case-sensitive. Description Example sin() Returns the sine of the sin(ch("TagNo.00001")) cos() Returns the cosine of the cos(ch("TagNo.00001")) tan() Returns the tangent of the tan(ch("TagNo.00001")) asin() Inverse sine asin(ch("TagNo.00001")) acos() Inverse cosine acos(ch("TagNo.
2.5 Setting Computations (Setting the Computation Channels) Time Functions The table below shows the functions related to time. The functions are not case sensitive. Only integer numeric constants can be written in the function parameters , , , , and in the table below. The terminology used in the description in the table are defined below. Edge: Returns 1.0 for computation immediately after the specified absolute or relative time. Previous edge: Returns 1.
2.5 Setting Computations (Setting the Computation Channels) • Timer () 1 2 Integration Monitor Up to eight timers can be specified. The timer is specified using a number between 1 and 8. Select the timer operation from the choices below. Select Off to disable the timer. Edge: Returns 1.0 when performing computation after the ON time elapses after the beginning of each interval. Specify the interval for detecting the condition and the ON time in ms. Level: Returns 1.
2.5 Setting Computations (Setting the Computation Channels) Event Functions A function used to carry out a given operation (event). They are not case-sensitive. They are mainly used in conjunction with the conditional operator (?). The return values are indicated below. The event is actually executed after the computation is complete. • Execution successful: 1.0 • Execution failure: 0.
2.5 Setting Computations (Setting the Computation Channels) Notes on Computation Ch 1 : Ch 2 : Ch 3 : Ch 4 : Ch 5 : Computation : : Data at time t : Data near time t : NaN : Oldest data : NaN t Measured value used in the computation of Example 2: If the disconnection (disconnected after ) is detected, NaN is used as the measured value of Ch 3.
2.5 Setting Computations (Setting the Computation Channels) • The value of NaN is considered true for the &&, ||, ^^, and ! operators. • For all other computations Computation is performed taking NaN as a value. Thus, NaN is returned as the computed result. • The numeric value display on the Monitor screen displays INVALID.
2.
2.6 Displaying the Monitor and Setting the Display Procedure Displaying the Monitor Screen and Settings on the Monitor Screen 1. Click Monitor to show the Monitor screen. 2. Change the display settings according to the explanation is the figure below. If the channels of different interval groups or channels of computation groups are assigned to the same display group, a monitor is displayed for each interval group or computation group (see the figure below).
2.6 Displaying the Monitor and Setting the Display Procedure 1 Explanation Display Groups and Group Names 2 Integration Monitor On the Monitor screen, the channels are divided into groups and displayed using waveforms or numerical values. The measured/computed values can be divided into up to ten groups. Up to thirty-two channels can be registered to a single group. The channels that can be registered are those where the Monitor check boxes are selected in the Channel setup screen.
2.6 Displaying the Monitor and Setting the Display Procedure Maximum and Minimum Values of the Y-Axis and the Display Format of the Scale Values Specify the maximum and minimum values of the Y-axis scale. The scale values can also be displayed using logarithmic format as shown below. Select the exponential format Selecting the Display Zone of the Waveform You can select from the following.
2.6 Displaying the Monitor and Setting the Display Procedure • Slide zone Displays the waveforms by slightly offsetting the display position of each waveform vertically regardless of the Zone settings on the Display setup screen. A single Y-axis scale of the active waveform can be displayed.
2.6 Displaying the Monitor and Setting the Display Procedure Trip Point You can display a trip line to indicate a particular value of interest (trip point) in the waveform display area. Two trip points (trip 1 is red, trip 2 is blue) can be set on each waveform. The trip line of the waveform corresponding to the left-most Y-axis is shown in the waveform display area.
2.6 Displaying the Monitor and Setting the Display Procedure Adding Marks Mark setup dialog box Mark Text of the specified mark App Note Marks cannot be placed when the monitor update is paused. Index Numeric Display You can switch from the waveform display to numeric display as shown below.
2.7 Starting and Stopping the Recording of Measured/Computed Data Procedure Record Start Operation Click the Record button on the Monitor screen, or choose Record from the Acquisition menu. or Choose Click When recording starts, the display changes as shown in the figure below. To divide the file during recording, click the Manual Save button or choose Manual Save from the Acquisition menu.
2.7 Starting and Stopping the Recording of Measured/Computed Data 1 When you click the Stop button on the Monitor screen or choose Stop from the Acquisition menu, a dialog box shown below opens. 2 Integration Monitor Click OK. When you click OK, the word “Saving”appears below the Stop button until the recording stops. 3 4 Explanation Start/Stop Recording Data is recorded according to the Start condition and Stop condition specified on the Acquisition setup screen.
2.8 Pausing the Updating of the Monitor Display and Reading Measured/Computed Values Using Cursors Procedure Pausing and Resuming the Updating of the Monitor Display To pause the monitor display, click the Pause button on the Monitor screen or choose Pause from the Acquisition menu. When paused, the word “Pause”under the button toggles between red and black. To resume the monitor display, click the Pause button on the Monitor screen or choose Pause release from the Acquisition menu.
2.8 Pausing the Updating of the Monitor Display and Reading Measured/Computed Values Using Cursors 1 To clear the cursors, choose Hide Cursor from the View menu. 2 Integration Monitor Choose 3 4 Explanation Pausing the Updating of the Waveform Display While monitoring the waveform, you can pause the updating of the waveform display and check the past waveforms (up to 1800 points). Pausing the display on the numeric display holds the current values immediately before the pause operation on display.
2.9 Setting Alarms and Alarm Indications Procedure Setting the Alarm For computation channels Click here Alarm 1 type Alarm value of alarm 1 Hysteresis of alarm 1 Alarm 2 type Alarm value of alarm 2 Hysteresis of alarm 2 Scroll to the right to display alarms 1 to 4 Rate-of-change interval when the Type is set to rHigh and rLow 1. Click Channel to show the Channel setup screen. 2. Click the Meas. Channel tab or the Math Channel tab. 3.
2.9 Setting Alarms and Alarm Indications 1 Explanation Alarm Types Alarm level settings and operation differ depending on the hardware style number of the MX100. See the table below. Alarm level setting (alarm 3 to alarm 4) MX100 Hardware Style Number* S3 S2 S1 Yes No No 3 Yes: Setting and operation allowed, No: Setting and operation not allowed * The style number is printed on the MX100 main module name plate. There are six types of alarms.
2.9 Setting Alarms and Alarm Indications Alarm Hysteresis You can set a width (hysteresis) to the values used to activate and release alarms. Alarm hysteresis prevents frequent activation and release of alarms when the measured/ computed value is unstable around the alarm value.
2.10 Digital Output Settings 1 Procedure 2 Alarm reference channel Select output or not output Select the alarm Turn ON/OFF Alarm 1 Turn ON/OFF Alarm 2 Turn ON/OFF Alarm 3 Turn ON/OFF Alarm 4 Select energize/de-energize of output relay Select hold/non-hold of output relay Integration Monitor Setting the Digital Output of Alarms 3 4 5 App Index 1. Click Channel to show the Channel setup screen. 2. Click the DO Channel tab. 3. Select the Use check boxes for the DO channels you wish to output. 4.
2.10 Digital Output Settings Explanation Alarm Reference Channel Range The alarm detection of multiple measurement channels or computation channels can be assigned to a single DO channel. If an alarm is detected on any one of the assigned channels, an alarm is output. A range of individual or consecutive channel numbers can be assigned. Note If connection with the PC is dropped, the state at the point at which the DO output was dropped is retained.
2.10 Digital Output Settings Energized/De-energized Operation of Output Relays 1 2 Integration Monitor You can select whether the output relay is energized or de-energized when an output event (such as an alarm) occurs. If De-energized is selected, the output relay behaves in the same fashion as when an output event occurs if the power is cut off. The default setting is Energized.
2.11 Analog/PWM Output Settings Procedure Analog Output Settings Select the output Selection of output Minimum/Maximum value action upon startup action upon errors of output span Select output Transmission output Output range Select the source action Sets the output value upon or no output reference channel selection Click here startup or error occurrence 1. Click Channel to show the Channel setup screen. 2. Click the AO Channel tab. 3. Select the Use check boxes for the AO channels you wish to output. 4.
2.11 Analog/PWM Output Settings 1 PWM Output Settings Sets the output value upon startup or error occurrence Select the resolution of the pulse output Enter the pulse interval coefficient 2 Integration Monitor Select output or no output Selection of output action upon errors Selection of output Minimum/Maximum value action upon errors of output span Transmission output Selection of output reference channel value upon startup 3 4 5 App Index 1. Click Channel to show the Channel setup screen. 2.
2.11 Analog/PWM Output Settings Turning Transmission Output ON and OFF Collectively for Analog Output/PWM Output 1. From the Action menu, choose Transmission output. Transmission output on all active channels on the AO and PWM channel tabs is started (All ON). 2. From the Action menu, choose Transmission output again. Transmission output on all active channels on the AO and PWM channel tabs is held at the values that were in use immediately before this operation was carried out.
2.11 Analog/PWM Output Settings Setting the User Output of the Pulse (PWM Channel) Drag the pointer to set Click the input box, and enter a value 1 2 Integration Monitor 3 Click to switch the display Overview of Output Action When Setting Holding of Previous Value of User Output 4 5 Value held Output value Power ON Measurement start Measurement stop App Measurement restarts Note If connection with the PC is dropped, the state at the point at which the AO or PWM output was dropped is retained.
2.12 Saving/Loading and Printing Setup Data (Project) Procedure Saving Setup Data (Project) 1. From the File menu, choose Save or Save As. 2. If you select Save As, enter the file name in the Save As dialog box and then click the Save button. If you attempt to save the file using an existing file name, a dialog box appears for you to confirm whether the file is to be overwritten. or Loading Setup Data (Project) 1. From the File menu, choose Open. 2.
2.12 Saving/Loading and Printing Setup Data (Project) 1 Printing Setup Data 1. From the File menu, choose Print. The Print dialog box opens. 2. In the Print dialog box, select the printer, print range, and the number of copies, Integration Monitor and then click OK. 2 3 Display example of the Print dialog box 4 5 App Print Preview From the File menu, choose Print Preview. The print image is displayed. Index Setting Up the Printer 1. From the File menu, choose Print Setup.
2.12 Saving/Loading and Printing Setup Data (Project) Explanation Save Destination The default save destination varies depending on the system that is running. File Name The extension is .mxp. The default file name is NewProject.mxp. Creating a New Project Create a new project when you wish to monitor the measured/computed data using new settings. Printed Contents of the Setup Data The following figure show how the setup data is printed.
2.13 Other Displays and Settings 1 Procedure 2 Integration Monitor Displaying the Version Information 3 1. From the Help menu, choose About. A dialog box containing version information opens.
Chapter 3 3.1 Viewer 1 Loading Data Files 2 Procedure 1. Start the Viewer. 2. From the File menu, choose Open. You can also click the Viewer The Open dialog box opens. 3 button on the toolbar. 4 5 App Index File information This information is also printed when the display data is printed 3. Select the file you wish to load and click Open. The waveform display window opens.
3.1 Loading Data Files Note If you open a file (file created by recording the data of multiple units (MX100s)) created using MXLOGGER (software sold separately), a confirmation dialog box opens with a message “This file is not synchronized. Synchronize data file?” If you click No, a Select Unit No. dialog box opens. Click the desired unit number button to display the measured data of the corresponding unit and the computed data of MXLOGGER.
3.1 Loading Data Files Synchronization The Integration Monitor of the MX100 Standard Software sends PC’s time information to the MX100 at measurement intervals. The PC’s time information is received along with the measured data when the data is received from the MX100. When measured data is recorded, the PC time information is also recorded. 1 2 3 Viewer PC Packet 4 PC Time 5 Packet MX100 PC Time Measured data ... ...
3.1 Loading Data Files The following limitations exist in the joining of files through synchronization. • If any of the divided files are missing, the files after (or before) the missing file are not joined. • If the total number of measured data points in the divided files exceeds 5 million, only files that would not cause the total number of displayed data points to exceed 5 million are joined.
3.1 Loading Data Files • • 1 2 3 Viewer • Even when synchronization is performed, the files before synchronization (.mxd extension) are not deleted. If you change the PC’s time while data is being monitored on the Integration Monitor of the MX100 Standard Software, the error in the clock counter value per millisecond of the PC used in the synchronization process will become large.
3.2 Display Settings Procedure Changing the Display on the Waveform Display Window Change the display settings according to the explanation is the figure below.
3.2 Display Settings 1 Changing the Display Using the Menu Choose appropriate commands from the Edit, View, and Window menus to change the display.
3.2 Display Settings Explanation Display Groups and Group Names The values of each channel that are loaded are divided into groups that were used during recording and displayed using waveforms or numeric values. The measured/computed values can be divided into up to fifty groups. Up to thirty-two channels can be registered to a single group. If you click the channel selection button on the General Display Settings window, the Channel No. dialog box opens (see the figure below).
3.2 Display Settings Maximum and Minimum Values of the Y-Axis and the Display Format of the Y-Axis Values The maximum and minimum values of the scale on the waveform display can be changed. If you click the Scale Calc. button on the General Display Settings window, the maximum and minimum values of the Y-axes on the selected channels are calculated automatically according to the maximum and minimum values of the data. The scale values can also be displayed using logarithmic format as shown below.
3.2 Display Settings • Full zone Displays all the waveforms over the full zone of the waveform display area regardless of the Zone settings on the General Display Settings window. A single Y-axis scale of the active channel can be displayed. Select full zone Waveform display area (full zone) • Slide zone Displays the waveforms by slightly offsetting the display position of each waveform vertically regardless of the Zone settings on the General Display Settings window.
3.2 Display Settings • Multi-axes zone All the specified Y-axis of the displayed waveforms are aligned horizontally. To hide a Y-axis, clear the Y-axes check box on the General Display Settings window. The display position is set using Zone on the General Display Settings window. Dragging the channel bar to the waveform display area shows the Y-axis of the corresponding channel. Dragging the Y-axis to the zone display area hides the Y-axis of the corresponding channel.
3.2 Display Settings Channel Color Clicking the channel color displayed in the Color column on the Display setup screen opens the Color dialog box. You can select the color of each channel using the Color dialog box. To create custom colors, click the Define Custom Colors button in the Color dialog box. Clipping of Waveforms By default (clip OFF), the waveform is not displayed when the measured/computed value exceeds the minimum/maximum value of the scale (see the lower left figure).
3.2 Display Settings Adding Marks Add mark icon Display the cursor Left end Flag Center 1 2 3 Viewer In addition to the marks placed using other programs such as the Integration Monitor, you can add marks at the positions where you click (displays a cursor) on the waveform display area. You can enter a text to be attached to the mark (“Mark”by default) using up to fifteen characters. You can also select Left, Center (default), Right, or Flag (small mark without text) for the Type.
3.2 Display Settings Showing/Hiding Alarms You can hide the alarm display area that indicates the status of alarm occurrence. The alarms are displayed in the same fashion as the alarms displayed in the waveform display area of the Integration Monitor. Show/Hide the alarms No alarm display Alarm display area Alarm/Mark List Display Displays a detailed list of alarms and marks. From the Window menu, choose Alarm/Mark List. You can also click the button on the toolbar.
3.2 Display Settings The alarm/mark list display’s cursors are linked with those of the waveform display and numeric display layout screens. Alarms or marks selected with the cursor in the alarm/ mark list display can be copied to the clipboard using the Edit > Copy command. The alarm/mark list can be converted to ASCII, Excel, or Lotus format (see section 3.7). Alarm display limitation: One file can display a maximum of 10000 alarms.
3.2 Display Settings Numeric Display Numeric display can be shown along with the waveform display window. When a group contain channels with different measurement intervals, the screen is divided accordingly. The display group, active channel, and mark display are synchronized to the waveform display. If cursors are displayed on the waveform display window, the data values between the cursors are displayed in red. Also, the data column corresponding to the cursor position is displayed in gray.
3.2 Display Settings 1 Initializing, Copying, and Pasting of Settings on the General Display Settings Window On the General Display Settings window, you can reset the settings to default or copy and pate the settings of one or multiple waveforms to the settings of other waveforms. You can copy and paste according to the procedure below. Drag the copy source waveform numbers (No. column), click the Copy button, drag the copy destination waveform numbers, and click the Paste button.
3.3 Reading Values Using Cursors Procedure 1. On the waveform display window, click the tag of the group from which you wish to read the value using cursors. 2. Click the position of the desired data in the waveform display area of the waveform display window. If you wish to read another point simultaneously, drag the cursor. Cursor A appears at the position where you first clicked; Cursor B appears at the position where you released the mouse button.
3.3 Reading Values Using Cursors 1 Clearing the Cursor From the Edit menu, choose Erase Cursor. The cursors are cleared as well as the cursor values displayed in the Cursor Value dialog box. Explanation 2 3 Reading Measured/Computed Values Using Cursors Viewer Two cursors can be displayed. The following values can be read on the Cursor Value window. • Values at the cursors. • Difference in the value between the cursors. • Absolute time and data number at the cursor position.
3.4 Statistical Computation over an Area of Measured/Computed Data Procedure 1. On the waveform display window, click the tab of the group on which you wish to perform statistical computation over an area. 2. Click the start position of the computation area in the waveform display area. A light-blue cursor appears in the waveform display area. If multiple waveform display areas are displayed, the cursor is displayed at the time position each waveform display area. 3.
3.4 Statistical Computation over an Area of Measured/Computed Data 1 Explanation Statistical Computation over an Area Specify using two cursors the interval over which computation is to be performed. If the cursors are not displayed, all the data are used in the statistical computation. The statistical parameters are the minimum value, the maximum value, the P-P value (maximum – minimum), the average value, and the rms value.
3.5 Saving the Display Settings Procedure 1. From the File menu, choose Save Display Setting File. You can also click the button on the toolbar. The display setting file is created in the same folder as the data files. When Closing the Viewer If you open a file and change the display settings, the dialog box below opens when you attempt to close the Viewer. To save the changed display settings, click Yes.
3.6 1 Saving the Display Template 2 Procedure Saving Templates 3 1. From the File menu, choose Save Template. Viewer The currently displayed settings are saved as a template file to the same folder as the displayed data. 4 5 Using Templates 1. From the File menu, choose Use Template. If the currently displayed data is not in the display conditions file, it is displayed according to the setting information of the template file residing in the same folder.
3.7 Converting Data Formats Procedure 1. From the Convert menu, choose To ASCII, To Excel, or To Lotus. 2. After entering various settings in the dialog box that opens, click OK. The file is created at the specified destination. Waveform, Numeric, or Waveform/Numeric Display Enter the range of group numbers to be converted Select whether to set the range to be converted using groups or channels Select the range of channels to be converted Click to open the selection dialog box.
3.7 Converting Data Formats Converted Data Specify the range using group numbers or channel numbers. You can select the channel number by clicking in the dialog box as shown below. 1 2 3 Conversion Range and Step You can specify the range using data numbers or by using cursors. For information on specifying the range by using cursors, see “Procedure” in section 3.4.
3.8 Printing the Data Procedure Setting the Contents to Be Printed 1. From the File menu, choose Print Settings. 2. In the Print Setting dialog box, set Range, Color, and Print Group, and then click OK. Enter a comment in the Comment box as necessary. When printing numeric values, only a print range need be selected.
3.8 Printing the Data 1 Executing the Print Operation 1. From the File menu, choose Print. The Print dialog box opens. 2. In the Print dialog box, select the printer, print range, and the number of copies, 2 and then click OK. When printing waveforms, you cannot set the print range. 3 Display example of the Print dialog box Viewer 4 5 App Print Preview Index 1. From the File menu, choose Print Preview. The print image is displayed. Setting Up the Printer 1.
3.8 Printing the Data Explanation Printed Items To print waveforms, open the waveform display window; to print numeric values, open the numeric display window; to print alarm or mark lists, open the alarm/mark list display window. If multiple waveform display windows or numeric display windows are open, click the window you wish to print. Also, to print alarm/mark lists, cursor values, or statistics calculated over and area along with waveforms, display those windows as well.
3.
3.
Chapter 4 4.1 Calibrator 1 Connecting the MX100 Notes before Using the Calibrator • If the MX100 is connected using the Integration Monitor, drop the connection before connecting the MX100 using the Calibrator. • If you change the module configuration before connecting with the Calibrator, use the Integration Monitor to reconfigure the system (see section 2.2). Procedure 2 3 4 1. Start the Calibrator. Calibrator The Unit Information window opens. 5 2.
4.1 Connecting the MX100 When the connection is established, information about the unit’s module configuration is displayed. To disconnect the connected MX100 Information about the connected MX100 To illuminate the 7-segment LED of the connected MX100 to see which MX100 is connected Modules attached to the connected MX100 If the specified IP address or host name is not correct, or the MX100 is connected by another software program, the following dialog box opens.
4.1 Connecting the MX100 1 Setting a Password for Startup You can use the following procedure to set up the Calibrator so that it prompts the user to enter a password upon startup. 1. From the File menu, click Set Password. 2 The password setting dialog box opens. 3 4 There are no restrictions on the characters that can be used in the password. However, the number of characters used must be no more than thirty. 3. Enter the same password in the Confirm New Password box, then click OK.
4.2 Calibration Procedure Notes on Calibration • Do not perform other operations during calibration. Take extra care because if you perform another operation while calibration is in progress, the module may suffer a break down. • When connecting the plate with screw terminals (model 772080) to the 10-CH, Medium-Speed Universal Input module, note that the terminal arrangement is different than when connecting clamped terminals.
4.2 Calibration Procedure 2. On the calibration detail setup screen that appears, select the channel and range to be calibrated and click the >> button. The figure below is an example of the 4-CH, High-Speed Universal Input Module. Index Items to be calibrated You can also click to select Select the measurement range to be calibrated from the list To move to the calibration execution screen 1 The figure below is an example of the 30-CH, Medium-Speed DCV/TC/DI Input Module. Channels are not selected.
4.2 Calibration Procedure The figure below is an example of the 6-CH, Medium-Speed, Four-Wire RTD Resistance Input Module. Channels are not selected. Items to be calibrated You can also click to select Select the measurement range to be calibrated from the list To move to the calibration execution screen 1 The figure below is an example of the 4-CH, Medium-Speed Strain Input Module. Channels are not selected.
4.2 Calibration Procedure 3. After making the connections according to the explanation given on the calibration execution screen 1, click the Calibrate button. For the 8-CH Medium Speed Analog Output module, after entering settings according to the explanation given on the calibration execution screen 1, click the Calibrate button. The figure below is an example of the 4-CH, High-Speed Universal Input Module. Exp.
4.2 Calibration Procedure If the calibration fails, a message “Failed to calibrate.”appears. Note For a description of the accuracy of the connected input module, the calibration accuracy, and the environmental conditions for calibration, see the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).
4.2 Calibration Procedure 4. After making the connections according to the explanation given on the calibration execution screen 2, click the Calibrate button. The figure below is an example of the 4-CH, High-Speed Universal Input Module. Exp. Calibration execution screen 2 1 2 3 4 Calibrator 5 App Start calibration Index The figure below is an example of the 8-CH, Medium-Speed Analog Output Module. Exp. Calibration execution screen 2 Start calibration Click the Calibrate button.
4.2 Calibration Procedure Explanation Calibrated Parameters • 4-CH, High-Speed Universal Input Module (MX110-UNV-H04) Calibration is carried out on a channel-by-channel basis.
4.
Chapter 5 5.1 Troubleshooting 1 Troubleshooting 2 If servicing is necessary, or if the instrument is not operating correctly after performing the corrective actions below, contact your nearest YOKOGAWA dealer for repairs. The 7-segment LED does not illuminate. Message Corrective Action The power switch is not ON. The supply voltage is too low. Turn ON the power switch. Check whether the voltage is within the supply voltage rating range. The fuse is blown. Servicing required.
5.1 Troubleshooting The MX100 cannot be detected from the PC or cannot be detected with the Search button. Message Corrective Action The LINK LED does not turn ON. The cable is broken. The LINK LED does not turn ON. There is a problem with the hub. Replace the Ethernet cable. The LINK LED does not turn ON. There is a problem with the PC. The ACT LED does not turn ON. There is a problem in the connection between the hub and the MX100. The ACT LED does not turn ON. There is a problem with the PC.
5.1 Troubleshooting The connected input/output module is not detected. Message Corrective Action Module connection or module startup error. Attached the module while the power was ON. The PC software, main module, and input/output module style and release numbers do not conform to the rule. Turn OFF the power. Detach the input/output module once and attach it again. Check the release number of the PC software and the main module style number, then upgrade the style. [PC software release no.
5.1 Troubleshooting The measured value is not correct. Message Corrective Action/Description The input wiring is not correct. Check the input wiring. The measured value indicates +Over or –Over. Change to an appropriate setting. The measurement range setting and input range do not match. The temperature error is large or is unstable. Change to the correct setting. The TC type setting and the type actually connected are different. The temperature error is large or is unstable.
5.2 1 Error Messages and Their Corrective Actions Error Messages on the Integration Monitor Error Code Corrective Action/Description IP address is invalid. Subnet mask is invalid. Default gateway is invalid. Failed to update. Enter the correct IP address. Enter the correct subnet mask. Enter the correct gateway address. Attempted to change the IP address while the connection is established, or the MX100 you wish to configure is connected by another software program. Disconnect the MX100.
5.2 Error Messages and Their Corrective Actions Error Messages on the Viewer Error Code Message Corrective Action/Description E0004 E0212 E3111 File write error. Can’t read file. Channels of the specified recording interval do not exist. E3114 Sampling data number is over the Viewer display limit of 5 M. Not a data file. File open error. The file cannot be created for some reason. Target files cannot be loaded, or are deleted.
Appendix Appendix 1 1 Configuring/Removing the Firewall Check the following before attempting to configure or remove a firewall. • Check the internet connection status using the ping command. If the ping fails, check the cables, hub, IP address settings, and other factors to determine the cause. • Check that procedures performed on the software are correct. If the problem cannot be solved after taking the above measures, it is likely that there will be problems with the firewall settings.
Appendix 1 Configuring/Removing the Firewall Windows XP SP3 1. From the Start menu, click Control Panel. The Control Panel opens. 2. Double-click Windows firewall. The Windows Firewall dialog box appears. 3. If Off is selected, click Cancel to exit. If On is selected, clear the Don’t allow exceptions check box.
Appendix 1 Configuring/Removing the Firewall 1 4. Click the Exceptions tab, then click the Add Program button. The Add program dialog box is displayed. 2 3 4 5 App Appendix 5. Select MX100 Standard, then click the OK button. Index 6. Click the Exceptions tab. Confirm that the MX100 Standard item was added, then select the MX100 Standard check box and click the OK button.
Appendix 1 Configuring/Removing the Firewall Windows Vista 1. From the Start menu, choose Control Panel. The control panel opens. 2. Click Allow a program through Windows Firewall. The User Account Control dialog box appears. 3. Click Continue. The Windows Firewall Settings dialog box is displayed.
Appendix 1 Configuring/Removing the Firewall 1 4. Click the General tab. If Off is selected, click Cancel to exit. If On is selected, clear the Block all incoming connections check box. 2 3 4 5 App Appendix Index 5. Click the Exceptions tab, then click the Add Program button. The Add a Program dialog box is displayed. 6. Select MX100 Standard, then click the OK button.
Appendix 1 Configuring/Removing the Firewall 7. Click the Exceptions tab. Confirm that the MX100 Standard item was added, then select the MX100 Standard check box and click the OK button.
Appendix 1 Configuring/Removing the Firewall 1 On Windows 7 1. Click the control panel on the Windows Start menu. The Control Panel appears. 2 3 4 2. Click System and Security. 5 App Appendix Index 3. Click Windows Firewall. The Windows Firewall Settings dialog box is displayed.
Appendix 1 Configuring/Removing the Firewall 4. Click Advanced settings. 5. Click Windows Firewall Properties. 6. Click the tab of the profile appropriate for your operating environment. When the Firewall state is Off, click Cancel to exit. When the Firewall state is On, and Inbound connections or Outbound connections is Block, proceed to next step.
Appendix 1 Configuring/Removing the Firewall 1 7. Click Inbound Rules. 2 3 4 5 App Appendix 8. Select the Program check box. Click Next.
Appendix 1 Configuring/Removing the Firewall 9. Select the This program path check box. Select MXStandardE.exe from the software’s installation folder and click Next. 10. Select the Allow the connection check box. Click Next.
Appendix 1 Configuring/Removing the Firewall 1 11. Specify the profiles for which this rule applies. Click Next. 2 3 4 5 App Appendix Index 12. Specify the name of this rule, and click Finish. 13. Follow the same procedure as the one in step 7 and select MXCalibE.exe and MxEngine.exe. 14. For Outbound Rules as well, follow the same procedure as the one in step 7 and select MXStandardE.exe, MXCalibE.exe and MxEngine.exe.
Index Index 1 2 Symbols E ±Overrange.......................................................................... 2-33 edit zone....................................................................... 2-46, 3-9 energize............................................................................... 2-59 equation...................................................................... 2-30, 2-31 error messages.....................................................................
Index M S MAC address......................................................................... 2-8 machine name....................................................................... 2-7 manual DO.......................................................................... 2-58 mark list............................................................................... 3-14 marks.......................................................................... 2-49, 3-13 measurement channels............................