User Manual

<3. Parameters>
3-15
IM 11M12D01-51E
3.4 Indication Check
Here we explain how to excute a Semi-Auto Indication check using HART communications. This
requires the AV550G Indication check mode to have been set to Semi-Auto or Auto.
For further details, refer to “Section 10. Indication Checkin the AV550G Instruction Manual.
Indication-Check-related parameters can not be set remotely using HART
communication.
Indication Check can not be canceled using HART communication.
Call up Indication checkdisplay.
1. Device setup ---> 2. Diag/Service ---> 3. Indication check
NOTE
1
2
3
(OK)
F030401-1.ai
F4
STU
1
AV550G:
Select channel.
1 Ch1
2 Ch2
3 Ch3
4 Ch4
5 Ch5
ABORT
ENTER
AV550G:
ABORT
ON
You select Ch1.
Do you want to
execute semi-auto
indication check?
AV550G:
Diag/Service
1 Loop test
2 Calibration
3 Indication check
4 Blow back
5 Calib log
HELP
SAVE
HOME
Select Channel for Indication Check
(Ch1 in this example).
Confirm that you want to execute
Indication Check
Indication Check is started using
parameters (Check Time, Hold Time)
set in the AV550G
Display
Operation