User Manual

<3. Parameters>
3-14
IM 11M12D01-51E
3.3 Calibration
Here we explain how to excute Semi-Auto Calibration using HART communications.
This requires the AV550G Calibration mode to have been set to Semi-Auto or Auto.
For further details, refer to “Section 9. Calibrationin the AV550G Instruction Manual.
Calibration-related parameters can not be set remotely using HART communication.
Calibration can not be canceled using HART communication.
Call up Calibrationdisplay.
1. Device setup ---> 2. Diag/Service ---> 3. Calibration
NOTE
1
2
3
(OK)
F030301-1.ai
F4
STU
1
AV550G:
1 Ch1
Select channel.
2 Ch2
3 Ch3
4 Ch4
5 Ch5
ABORT
ENTER
AV550G:
ABORT
ON
You select Ch1.
Do you want to
execute semi-auto
calibration?
AV550G:
Diag/Service
1 Loop test
2 Calibration
3 Indication check
4 Blow back
5 Calib log
HELP SAVE
HOME
Select Channel to be calibrated
(Ch1 in this example).
Confirm that you want to execute
calibration
Calibration is started using parameters
(Cal. Time, Hold Time) set in the
AV550G
Display
Operation