User Manual
<3. Parameters>
3-8
IM 11M12D01-51E
3.2.2  Multidrop Mode
Field devices in multidrop mode refer to the connection of several eld devices on a single 
communication line. Up to 15 eld devices can be connected when set in multidrop mode. To activate 
multidrop communication, each eld device address must be changed to a unique number in the 
range 1 to 15. This change deactivates the 4 to 20mA output and changes it to 4mA.
Setting of Multidrop Mode
Call up “Poll addr” display.
1.Device setup ---> 4.Detailed setup ---> 4.Output condition ---> 1.HART output  
--->1.Poll addr
Call up “Auto Poll” display.
HART output ---> Online Menu ---> Main Menu ---> 4.Utility ---> 1.Congure Communication  
---> 1.Polling
"
1
"
 (ENTER) (SEND)
Set the polling address (a
number from 1 to 15) and
press 
ENTER (F4)
.
Then Press
 SEND (F2).
F4
F2
F030210-1.ai
AV550G:
Poll addr
 0
 0 
ESC
DELHELP ENTER
Display
Operation
1
(HOME)
Return 
" Online Menu " 
 with
HOME (F3)
.
2
Return to 
" Main Menu " 
 with
a 
" previous" 
 key.
3
Select
" Utility " 
.
F3
F030211-1.ai
JLK
4
AV550G:
HART output
 1 Poll addr   1
 2 Num req preams   5
 3 Burst mode   On
 4 Burst option   PV
HOME
SENDHELP
AV550G:
1
Device setup
Online
2 PV
3 PV URV
4 PV LRV
21.64
21
0
%
%
%
SAVE
HART Communicator
 1 Offline
 2 Online
 3 Frequency Device
 4 Utility
Display
Operation