3.2 Setting Parameters
3.2.1 Burst Mode
The AV550G continuously sends the data stored in it when burst mode is set “On”. Either one of
measured value, % output value, or 4 to 20 mA output value can be selected and sent. The data
is sent periodically at 75 ms intervals as a digital signal when the AV550G is set in burst mode.
Therefore, communication by the HART simultaneous communicator is also possible.
Setting of Burst Mode
Call up “Burst option” display.
1.Device setup ---> 4.Detailed setup ---> 4.Output condition ---> 1.HART output --->
4.Burst option
Display Operation
(ENTER)
Set data to be sent.
Measured value (PV)
% output value
(% range/current)
4 to 20 mA output value
(Process vars /crnt)
Call up
" Burst mode"
display.
3.HART output 3.Burst mode
1
(ENTER)
Set
"
On
"
and press
ENTER
(F4).
2
(SEND)
Press
SEND (F2).
F4
F4
F2
F030209-1.ai
AV550G:
Burst option
PV
% range/current
Process vars/crnt
ENTER
ESCHELP
AV550G:
Burst mode
Off
ENTER
ESC
off
on
AV550G:
HART output
1 Poll addr 0
2 Num req preams 5
3 Burst mode On
4 Burst option PV
HOME
SENDHELP
********