10. Other Functions
Table 10.8 Initialization Parameters and Defaults - part 2
T10.8E.eps
During warmup OFF
During output range switching OFF
During maintenance OFF
During calibration ON
During indication check OFF
During blowback OFF
3rd check gas OFF
Function No function (disabled)
Action Normally closed
Item Initialization Parameter Default
Alarms
Hysteresis 0.1% O2
Alarm contact action delay 3 sec
High High Alarm OFF
Alarm setting 100.0% O
2
High Alarm OFF
Alarm setting 100.0% O
2
Low Alarm OFF
Alarm setting 0.0% O
2
Low Low Alarm OFF
Alarm setting 0.0% O
2
Instrument internal temperature alarm log Deleted
Removal of channel in alarm condition NO
Contacts
Contact output 1
Alarms All OFF
Other settings
During warmup ON
During output range switching OFF
During maintenance OFF
During calibration OFF
During indication check OFF
During blowback OFF
3rd check gas OFF
Relay state when powered Energized
Contact output 2
Alarms All OFF
Other settings
Relay state when powered Energized
Contact output 3
Alarms
High High Alarm OFF
High Alarm ON
Low Alarm ON
Low Low Alarm OFF
Calibration coefficient alarm OFF
Calibration gas pressure low OFF
Error OFF
Instrument internal temperature alarm OFF
Other settings All OFF
Relay state when powered Energized
Contact output 4
Alarms High High Alarm OFF
High Alarm OFF
Low Alarm OFF
Low Low Alarm OFF
Calibration coefficient alarm OFF
Calibration gas pressure low OFF
Error OFF
Instrument internal temperature alarm OFF
Other settings All OFF
Relay state when powered Energized
Contact input 1
Contact input 2
Alarms