Manual
IM 11M12D01-01E
10-28
Table 10.7 Initialization Parameters and Defaults -part1
T10.7E.eps
Mode Manual
Points Span - Zero
Zero gas concentration 1.00% O
2
Span gas concentration 21.00% O
2
Minimum 0% O
2
Maximum 25% O
2
Output smoothing time constant 0 sec (disabled)
Output mode Linear
During warmup 4 mA
Preset value 4 mA
Under maintenance Last measured value
Preset value 4 mA
Under calibration, Indication check, During Blowback Last measured value
Preset value 4 mA
Error Last measured value
Preset value 4 mA
Group invalid 4 mA
Preset value 4mA
Process alarm Last measured value
Preset value 4mA
Item Initialization Parameter Default
Display
Tag name Deleted
Trend Graph
Displayed channels All ON
Sampling interval 30 sec
Graph upper limit 25.0% O
2
Graph lower limit 0.0% O
2
Auto Return Time 0 min. (disabled)
Display contrast Medium
Display Auto power off 0 min (disabled)
Display language Not affected
CAL Indication Check
Hold time 3 min. 00 sec
Calibration / Indication check time 3 min . 00 sec.
Calibration / Indication check interval 1 day 00 hrs.
Start date (day/month/year) 01 / 01 / 01
Start time 00 : 00
Blowback
Mode No function (disabled)
Hold time 3 min 00 sec
Blowback time 10 min 00 sec
Blowback interval 30 days 00 hours
Start date (day/month/year) 01 / 01 / 01
Start time 00 : 00.
mA - output
Output range
Output holod setting