Manual
IM 11M12D01-01E
9-9
9. Calibration
Table 9.1 When Auto Calibration conflicts with Other Commands
T9.1E.eps
 X : Cannot perform calibration 
Wait : Wait until other command finishes before starting calibration 
Purging 
the 
channel
Other 
channel 
CAL
Indication 
Check in 
other channel 
Blowback in 
other 
channel
CAL chan.
in warmup
status
CAL chan.
in ERR
status
CAL chan.
now 
disabled
Maintenance
Man. Cal.
SemiAuto
AutoCal
Wait Wait Wait Wait Wait
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
xxx
Other
commands
<Shipping-time defaults>
At shipping-time (or after resetting parameters) default values are as shown in Table 9.2
below.
Table 9.2
T9.2E.eps
Item Initial value
Mode Manual
Points Span-Zero
Zero gas conc.  1.00%
Span gas conc.  21.00%
Hold time  3 min. 00 sec.
Cal. time 3 min. 00 sec.
Interval 1 days 00 hrs
Starting date yy01mm01dd01
Starting time hh00mm00 