Manual
IM 11M12D01-01E
8-15
8. Detailed Data Setting
8.4 Assigning Contact Inputs
External contact inputs allow remote operation of the following functions shown in
Table 8.6 .
Table 8.6 Contact Input Functions
T8.6E.eps
Output Range Switching While contact signal is applied, output range 2 is selected 
and [Range] is displayed on the display. See Fig. 8.15. 
CAL start Contact input starts Semi-Automatic Calibration. 
Calibration Mode setting must be Semi-Auto or [Auto]. 
Contact signal must be applied for at least 1 sec. Even 
if input signal continues to be applied, calibration is not 
repeated unless contact input is released then reapplied. 
Indication Check Start  Contact input starts Semi-Automatic Indication Check. 
Indication Check setting must be Semi-Auto or [Auto]. Contact 
signal must be applied for at least 1 sec. Even if input signal 
continues to be applied, Indication Check is not repeated 
unless contact input is released then reapplied. 
Blowback start Contact input starts Blowback. Contact signal must be 
applied for at least 1 sec. Even if input signal continues to 
be applied, Blowback is not repeated unless contact input 
is released then reapplied.
CAL gas pressure low Contact input disables Semi-Auto or Auto Calibration. 
Process Gas Alarm  While contact signal is input, for safety reasons heater power 
is turned off. When contact input is released, heater power is 
reapplied and warmup begins. 
When warmup ends then Measuring mode starts 
Item Function
Tag:
Ch1
 21.0%
Ch2
 21.0%
Ch3
 21.0%
Ch4
 21.0%
Ch5
 21.0%
Ch6
 21.0%
Ch7
 21.0%
Ch8
 20.7%
F8.15.EPS
Range
Fig. 8.15 During Range Switching