Manual
IM 11M12D01-01E
8-7
8. Detailed Data Setting
<Parameter Setting Procedure>
1) From "mA-output setup" on the Commissioning display, select "mA-outputs hold"
and touch the Enter key. A display like Fig. 8.9 is displayed.
2) Select an item you want to change, and touch the Enter key.
3) In the selection window that appears, select the desired Hold method and touch the
Enter key.
4) Select other items as desired and set as desired, as in the previous step. If you select
Preset Value then you can set a current output value.
5) Touch the Reject key to return to the "mA-outputs hold" display.
6) If you select Preset Value and touch the Enter key then the display shown in Fig. 8.10
appears.
7) Selecting a preset setting and touching the Enter key displays the numeric entry
display, allowing you to enter a preset current in the range 2.4 mA to 21.6 mA.
Enter
F8.9E.EPS
r Maintenance: Non-Hold
r Cal/Blow/Ind: Hold
r Proc. gas alm: Hold
r Error: Hold
r Ave group invalid:
4mA
mA-outputs hold
Warm up: 4mA
Enter
F40E.EPS
Warm up: 4. 0 mA
r Maintenance: 4. 0 mA
r Cal/Blow/Ind: 4. 0 mA
r Proc. gas alm: 4. 0 mA
r Error: 4. 0 mA
r Ave group invalid:
4. 0 mA
Preset vaiue
Fig. 8.9 Current output hold Fig. 8.10 Preset value setting
<Default setting>
Output hold settings are set to defaults in Table 8.2 prior factory shipment or after data
initialization..
Tbale 8.2 Defaults for output hold and preset values
Status Output hold value Preset value
Warmup, Purging 4mA 4mA
(Under) Maintenance Hold last value 4mA
(Under) Calibration
Indication Check
Blowback
Process Gas Alarm Hold last value 4mA
Error Hold last value 4mA
No Valid Channels in Averaging Group 4mA 4mA
Hold last value 4mA
T8.2Eeps