8. Detailed Data Setting
<Parameter Setting Procedure>
1) From "mA-output setup" on the Commissioning display, select "mA-outputs hold"
and touch the Enter key. A display like Fig. 8.9 is displayed.
2) Select an item you want to change, and touch the Enter key.
3) In the selection window that appears, select the desired Hold method and touch the
Enter key.
4) Select other items as desired and set as desired, as in the previous step. If you select
Preset Value then you can set a current output value.
5) Touch the Reject key to return to the "mA-outputs hold" display.
6) If you select Preset Value and touch the Enter key then the display shown in Fig. 8.10
appears.
7) Selecting a preset setting and touching the Enter key displays the numeric entry
display, allowing you to enter a preset current in the range 2.4 mA to 21.6 mA.
Enter
F8.9E.EPS
r Maintenance: Non-Hold
r Cal/Blow/Ind: Hold
r Proc. gas alm: Hold
r Error: Hold
r Ave group invalid:
 4mA
mA-outputs hold
 Warm up: 4mA
Enter
F40E.EPS
 Warm up: 4. 0 mA
r Maintenance: 4. 0 mA
r Cal/Blow/Ind: 4. 0 mA
r Proc. gas alm: 4. 0 mA
r Error: 4. 0 mA
r Ave group invalid:
 4. 0 mA
Preset vaiue
Fig. 8.9 Current output hold Fig. 8.10 Preset value setting
<Default setting>
Output hold settings are set to defaults in Table 8.2 prior factory shipment or after data
initialization..
Tbale 8.2 Defaults for output hold and preset values
Status  Output hold value Preset value
Warmup, Purging   4mA  4mA
(Under) Maintenance Hold last value   4mA
(Under) Calibration 
Indication Check 
Blowback 
Process Gas Alarm  Hold last value   4mA
Error Hold last value   4mA
No Valid Channels in Averaging Group  4mA   4mA
Hold last value   4mA