7. Startup
7.5 Supply Power to Averaging Converter
CAUTION
To avoid temperature changes around the sensor, it is recommended that (rather than
turning it on and off) power be continuously supplied to the Oxygen Analyzer if it is
used in an application where it is used periodically.
It is also recommended to flow a span gas (instrument air) beforehand.
Before connecting power check that power switch is set to "O (Off)". After connecting
power. Turn power switch to "I (On)". The startup display is displayed for about 10
seconds then the multi-channel cell temperature display shown in Fig. 14 appears.
Confirm that the displayed temperatures of each channel gradually increase.
After the cell temperatures reach their target setting of 7508C and stabilize, the display
switches to show oxygen concentration. When all the channels have switched to display
oxygen concentration, the red [Warmup] and [Hold] alarm marks at the bottom of the
display disappear.
Warmup
Tag:
Ch1
 5148C
Ch2
 5158C
Ch3
 5138C
Ch4
 5148C
Ch5
 5148C
Ch6
 5148C
Ch7
 5138C
Ch8
 5198C
Hold
F7.4E.EPS
Fig. 7.4 Display during warmup
Note
Depending on furnace conditions, the cell temperature may overshoot the target by
10~158C during the warmup phase. This is not an abnormality.