<APPENDIX 1. LIST OF PARAMETERS FOR EACH BLOCK OF AV550G>
A-11
IM 11M12D01-61E
176 2176 CH8_MIN_VALUE Denotes the minimum oxygen concentration of channel 8.
177 2177 CH8_MAX_VALUE Denotes the maximum oxygen concentration of channel 8.
178 2178 CH8_AVE_VALUE Denotes the average oxygen concentration of channel 8.
179 2179 CH8_MIN_DATE Denotes the date and time when the minimum oxygen
concentration of channel 8 was calculated.
180 2180 CH8_MAX_DATE Denotes the date and time when the maximum oxygen
concentration of channel 8 was calculated.
181 2181 CH1_SMART_CALIB_DA
TE
Denotes the recommended date of channel 1’s next
calibration.
182 2182 CH2_SMART_CALIB_DA
TE
Denotes the recommended date of channel 2’s next
calibration.
183 2183 CH3_SMART_CALIB_DA
TE
Denotes the recommended date of channel 3’s next
calibration.
184 2184 CH4_SMART_CALIB_DA
TE
Denotes the recommended date of channel 4’s next
calibration.
185 2185 CH5_SMART_CALIB_DA
TE
Denotes the recommended date of channel 5’s next
calibration.
Appendix1-3_6.ai
186 2186 CH6_SMART_CALIB_DA
TE
Denotes the recommended date of channel 6’s next
calibration.
187 2187 CH7_SMART_CALIB_DA
TE
Denotes the recommended date of channel 7’s next
calibration.
188 2188 CH8_SMART_CALIB_DA
TE
Denotes the recommended date of channel 8’s next
calibration.
189 2189 CH1_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop
calibrating
channel 1.
190 2190 CH2_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 2.
191 2191 CH3_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 3.
192 2192 CH4_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 4.
193 2193 CH5_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 5.
194 2194 CH6_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 6.
195 2195 CH7_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 7.
196 2196 CH8_SEMIAUTO_CAL_S
TART
1 (stop) AUTO Determines whether to start or stop calibrating
channel 8.
197 2197 CH1_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 1.
198 2198 CH2_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 2.
199 2199 CH3_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 3.
200 2200 CH4_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 4.
201 2201 CH5_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 5.
202 2202 CH6_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 6.
203 2203 CH7_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 7.
204 2204 CH8_INDICATION_STAR
T
1 (stop) AUTO Determines whether to start or stop checking the
indication of channel 8.
205 2205 BLOWBACK_START 1 (stop) AUTO Determines whether to start or stop blowing back.
206 2206 CAL_GAS_PRESS_DRO
P_SW
1 (off) AUTO Sets the on or off state for a drop in the calibration gas
pressure.
207 2207 PROCESS_GAS_ALARM
_SW
1 (off) AUTO Sets the on or off state for process gas alarms.
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto
semi-auto