Instruction Manual

<APPENDIX 1. LIST OF PARAMETERS FOR EACH BLOCK OF AV550G>
A-7
IM 11M12D01-61E
26 2026 ERROR_SW_VALUE_D Shows that an error has occurred in the AV550G.
27 2027 IN_UNIT 1342 O/S Denotes the unit of MAO data to be indicated on the AV550G’s
display.
28 2028 IN_DISPLAY_FORMAT 0 O/S Denotes the decimal places of MAO data to be indicated on
the AV550G’s display.
29 2029 USE_IN_NO 1 O/S Denotes the IN number of MAO data to be indicated on the
AV550G’s display.
30 2030 PV1_MIN_VALUE Denotes the minimum oxygen concentration of averaging A.
31 2031 PV1_MAX_VALUE Denotes the maximum oxygen concentration of averaging A.
32 2032 PV1_AVE_VALUE Denotes the average oxygen concentration of averaging A.
33 2033 PV1_MIN_DATE Denotes the date and time when the minimum oxygen
concentration of averaging A was calculated.
34 2034 PV1_MAX_DATE Denotes the date and time when the maximum oxygen
concentration of averaging A was calculated.
35 2035 PV2_MIN_VALUE Denotes the minimum oxygen concentration of averaging B.
36 2036 PV2_MAX_VALUE Denotes the maximum oxygen concentration of averaging B.
37 2037 PV2_AVE_VALUE Denotes the average oxygen concentration of averaging B.
Appendix1-3_2.ai
38 2038 PV2_MIN_DATE Denotes the date and time when the minimum oxygen
concentration of averaging B was calculated.
39 2039 PV2_MAX_DATE Denotes the date and time when the maximum oxygen
concentration of averaging B was calculated.
40 2040 PV3_MIN_VALUE Denotes the minimum oxygen concentration of averaging C.
41 2041 PV3_MAX_VALUE Denotes the maximum oxygen concentration of averaging C.
42 2042 PV3_AVE_VALUE Denotes the average oxygen concentration of averaging C.
43 2043 PV3_MIN_DATE Denotes the date and time when the minimum oxygen
concentration of averaging C was calculated.
44 2044 PV3_MAX_DATE Denotes the date and time when the maximum oxygen
concentration of averaging C was calculated.
45 2045 CH1_CELL_VOLT Denotes the cell electromotive force of the channel-1 detector.
46 2046 CH1_HEATER_TEMP Denotes the heater temperature of the channel-1 detector.
47 2047 CH1_CJ_TEMP Denotes the cold junction temperature of the channel-1
detector.
48 2048 CH1_TC_VOLT Denotes the thermocouple electromotive force of the
channel-1 detector.
49 2049 CH1_CJ_VOLT Denotes the cold junction voltage of the channel-1 detector.
50 2050 CH1_CELL_RESISTANC
E
Denotes the cell resistance of the channel-1 detector
measured during calibration.
51 2051 CH1_CJ_RESISTANCE Denotes the cold junction resistance of the channel-1 detector.
52 2052 CH1_ZERO_CAL_COEFF Denotes the zero-point correction factor of the channel-1
detector.
53 2053 CH1_SPAN_CAL_COEFF Denotes the span-point correction factor of the channel-1
detector.
54 2054 CH1_CELL_ROBUSTNES
S
Denotes the cell service life of the channel-1 detector.
55 2055 CH1_HEATER_ON_TIME Denotes the power-on rate of the channel-1 detector’s heater.
56 2056 CH1_RESPONSE_TIME Denotes the response time of the channel-1 detector
measured during calibration.
57 2057 CH1_MIN_VALUE Denotes the minimum oxygen concentration of channel 1.
58 2058 CH1_MAX_VALUE Denotes the maximum oxygen concentration of channel 1.
59 2059 CH1_AVE_VALUE Denotes the average oxygen concentration of channel 1.
60 2060 CH1_MIN_DATE Denotes the date and time when the minimum oxygen
concentration of channel 1 was calculated.
61 2061 CH1_MAX_DATE Denotes the date and time when the maximum oxygen
concentration of channel 1 was calculated.
62 2062 CH2_CELL_VOLT Denotes the cell electromotive force of the channel-2 detector.
63 2063 CH2_HEATER_TEMP Denotes the heater temperature of the channel-2 detector.
64 2064 CH2_CJ_TEMP Denotes the cold junction temperature of the channel-2
detector.