Manual

<8.Appendix>
8-1
IM 11M10B01-01E
8. Appendix
8.1 ParameterSettings
Table8.1 ParameterSettingList(CustomerSettings)
Menu Parameter 1 Parameter 2 Parameter3 Functiondetail Defaults Description
Customer
Settings
Reference
section
Output range
(4-20 mA DC)
“rnG
“Auto” Auto range “1 (Note 1)
6.1.1
“MAn Manual range 0-1000 ppm
“FrEE”
Partial range 0-1000 ppm
“Fr.Lo”
Low limit 0 ppm
“Fr.HI”
High limit 1000 ppm
Secondary output
“oUt2 “1”/“5”/“10
Voltage output “1 0-1V DC 6.1.2
Hold function “HoLd”
“non”/“PrEV
Previous or non non Non hold 6.1.3
Burnoutfunction nAMU
“non”/“Er.Lo”/“Er.HI”
Hi/Low/non non Non burnout 6.1.4
O
2conc.Hi/Low
alarm
“ALM
“oFF”
Non alarm
oFF Non alarm 6.1.5
“ALL”
“AL.Lo”
“AL.HI”
Low limit
Hi limit
“AL.HI”
Hi limit only
“AL.Lo”
Low limit only
Calibration gas
concentration
“SEt.C”
“1Pnt” 1 point 1000 ppm
Cal-gas conc.
(Note 2)
6.1.6
“2Pnt”
2 points
“Zero
Zero-gas conc. 100.0 ppm (Note 3)
“SPAn”
Span-gas conc. 1000 ppm (Note 4)
“ALL
3 points
“10
10 ppm gas conc. 10.00 ppm (Note 5)
“1000
1000 ppm gas conc. 1000 ppm (Note 6)
Outputsmoothing
“SMoo “0to60
Smoothing factor. 0 sec Non (Note 7) 6.1.8
Multi-selector
“MLS” “0”/“1”/“2”/“3”
Flow path No.1 to 3 “0 Non use 6.1.9
(Note 1) Auto Range between 0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-1%, 0-10%, and 0-100%.
(Note 2) Calibration gas concentration input range: 0.90 ppm to 12.00%, 35.00% to 100.0%. Gas concentration can also be set from “CAL” calibration.
(Note 3) “ZEro” input range: 1.00 ppm to 100.0%. Gas concentration can also be set from “CAL” calibration,
(Note 4) “SPAn” input range; 1.00 ppm to 100.0%. Input is not possible if the smallest span is less than 20% of the range after inputting “SPAn.Gas concentration can
also be set from “CALcalibration,
(Note 5) Calibration gas concentration input range: 8.00 to 20.00 ppm. Gas concentration can also be set from “CALcalibration,
(Note 6) Calibration gas concentration input range: 800 to 2000 ppm. Gas concentration can also be set from “CAL” calibration,
(Note 7) If the output smoothing is set, the smoothed value becomes an instantaneous value which is displayed and output in analog format. The smoothed value is also
used as the instantaneous value (measured value) for checking concentration alarms and RS232 communication.
(Note 8) Set parameters will not be erased by turning the power off and on.