<6.OperationandParameters>
6-4
IM 11M10B01-01E
6.1.2 SettingSecondaryOutput“oUt2
Operation:Holddownthe[MODE]keyfor2seconds,select“oUt2”withthe[▲],[▼]keys,and
pressthe[ENT]key.
With respect to the secondary output, select one of the three types of voltage output: 0-1 V is “1,0-5
Vis“5,”and0-10VDCis“10.”Figure6.3showsanexampleofsetting0-1VDC“1.”
CAL
rnG
oUt2
HoLd
▼↓↑▲
▼↓↑▲
▼↓↑▲
▼↓↑▲
[ENT]
1
oUt2
PPM
5
oUt2
10
oUt2
▼↓↑▲
▼↓↑▲
▼↓↑▲
▼↓↑▲
[ENT]
1
oUt2
PPM
[ENT]
MODE
(0-10 V D C )
(0-5 V D C )
%
%
To “rEV”
Hold down
for 2 seconds.
To “10”
0-1 V DC is confirmed.
Return to “oUt2.”
To “1”
Figure6.3 SettingSecondaryOutput“oUt2”
6.1.3 SettingHOLDFunction“HoLd”
Operation:Pressthe[MODE]keyfor2seconds,selectHoLdwiththe[▲],[▼]keys,andpress
the[ENT]key.
Set the output state in the maintenance mode. Select eithernon” or “PrEV.” The former option means
nohold,andthelattermeansholdthepreviousvalue.Thefollowingisanexampleofnohold.
HoLd
▼↓↑▲
CAL
▼↓↑▲
▼↓↑▲
[ENT]
HoLd
PPM
HoLd
▼↓↑▲
▼↓↑▲
▼↓↑▲
[ENT]
HoLd
PPM
[ENT]
MODE
non
PrEV
non
To “rEV”
Hold down
for 2 seconds.
To “PrEV”
No hold is confirmed.
Return to “HoLd.”
(Hold the previous value.)
To “non”
%
%
Figure6.4 SettingHOLD“HoLd”