Manual

<2.Specications>
2-6
IM 11M10B01-01E
Panelmounttypewithaspirator(OX400--A--/P)
40 to 50
(Line connection : in case of Rc1/4)
(Line connection : in case of Rc1/4)
(Line connection : in case of 1/4NPT)
(Line connection : in case of 1/4NPT)
Mount
Ventilation
holes
Bushing
Bushing
Aspirator
Aspirator
Connector
4xM5 screw
Panel Cutout
60 to 70
101.6
±0.3
135
+2
0
240
±1
225
±1
19.3330
20
10 to 20
FrameFrame
Screw
Seal
6 6
132
101.6
260
238
3. Make sure the panel is at least 2 mm thick.
2. Maintain at least 100 mm of free space around the measuring instrument in order to ensure
adequate ventilation.
1. Make sure the bottom supports do not block the ventilation outlet on the bottom panel of
the measuring instrument.
Notes on mounting
Unit: mm