Manual
<2.Specications>
2-5
IM 11M10B01-01E
Consumables
Item Partno.andrating Qt’y
Activatedlterelementkit(15timesreplacement) K9643KK 1
Filter kit (5 times replacement) K9643KL 1
Sensor assembly (including O-ring) K9643KG 1
Snap ring (retainer) Y9011EV 1 (*)
Plate K9213FB 1 (*)
Filter K9643FB 1 (*)
Snap ring plier K9643ZE 1
(*) Qt’y of 10 pieces or more can be purchased.
2.5 ExternalDimensions
Panelmounttypewithbuilt-inpumpornosuctiondevice(OX400----/P)
135
+2
0
101.6
±0.3
240
±1
225
±1
19.3330
20
10 to 20
6 6
132
101.6
260
238
Seal
Ventilation
holes
FrameFrame
Screw
4xM5 screw
Panel Cutout
Mount
Note: Pump ON/OFF switch; when no suction device [-N] is specified for the sampling method,
this switch is not installed.
3. Make sure the panel is at least 2 mm thick.
2. Maintain at least 100 mm of free space around the measuring instrument in order to ensure
adequate ventilation.
1. Make sure the bottom supports do not block the ventilation outlet on the bottom panel of
the measuring instrument.
Notes on mounting
Unit: mm