<2.Specications>
2-3
IM 11M10B01-01E
Line connection
Gasinlet:Rc1/4or1/4NPTfemale
Gasoutlet:Rc1/4or1/4NPTfemale
Electricalconnection
Externaloutputterminals:M3screw
Contactinput/outputterminals:M3screw
Serial communication: D-sub 9 pin connector
Ground: within power cord connector
Environmentandoperationalconditions
Installationconditions:Indoors,panelorwallmounting,nonexplosionarea
Ambient temperature: 0 to 40°C, non-condensing
Ambient humidity: 5 to 85% RH
Storage temperature: -5 to 50°C
Conformance to SafetyandEMCstandards
US
C
R
Safety : EN 61010-1
CAN/CSA-C22.2No.61010-1-04,ULStd.No.61010-1
EMC: EN 61326-1 Class A, Table 2 (for use in industrial locations)
EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
(Note1) This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
(Note1) The current or voltage signal output cable length and contact input cable length must be no
longer than 30 m for CE marking. RS232 connection cable length must be no longer than 3 m.
Installation altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note:Installationcategory,calledover-voltagecategory,speciesimpulsewithstandvoltage.
Category II is for electrical equipment.
Pollutiondegreeindicatesthedegreeofexistenceofsolid,liquid,gasorother
inclusions which may reduce dielectric strength. Pollution degree 2 is the normal indoor
environment.
2.2 Characteristics
Repeatability
: ±1% FS (Hereafter, either 10, 100, 1000 ppm, 1%, 10%, or 100% O
2
is FS)
Linearity
: ±2% FS
±3% FS (0-100 ppm or less)