<2.Specications>
2-3
IM 11M10B01-01E
Line connection
Gasinlet:Rc1/4or1/4NPTfemale
Gasoutlet:Rc1/4or1/4NPTfemale
Electricalconnection
Externaloutputterminals:M3screw
Contactinput/outputterminals:M3screw
Serial communication: D-sub 9 pin connector
Ground: within power cord connector
Environmentandoperationalconditions
Installationconditions:Indoors,panelorwallmounting,nonexplosionarea
Ambient temperature: 0 to 40°C, non-condensing
Ambient humidity: 5 to 85% RH
Storage temperature: -5 to 50°C
Conformance to SafetyandEMCstandards
US
C
R
Safety : EN 61010-1
CAN/CSA-C22.2No.61010-1-04,ULStd.No.61010-1
EMC: EN 61326-1 Class A, Table 2 (for use in industrial locations)
EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
(Note1) This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
(Note1) The current or voltage signal output cable length and contact input cable length must be no
longer than 30 m for CE marking. RS232 connection cable length must be no longer than 3 m.
Installation altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note:Installationcategory,calledover-voltagecategory,speciesimpulsewithstandvoltage.
Category II is for electrical equipment.
Pollutiondegreeindicatesthedegreeofexistenceofsolid,liquid,gasorother
inclusions which may reduce dielectric strength. Pollution degree 2 is the normal indoor
environment.
2.2 Characteristics
Repeatability
: ±1% FS (Hereafter, either 10, 100, 1000 ppm, 1%, 10%, or 100% O
2
is FS)
Linearity
: ±2% FS
±3% FS (0-100 ppm or less)