Manual
<2.Specications>
2-2
IM 11M10B01-01E
Contact output: 3 outputs
Error contact, O
2
concentrationhigh/lowalarmcontact,rangemarkeroutput
Multi selector (optional)
:Contactoutputforswitchingsamplegasow,measumentowinformationcontactoutput.
Note:Fordetails,seeexternaldimensions.
Contactoutputspecications
General : Relay
Nominal contact capacity:
2 A 30V DC, 2 A 240 V AC (120 V AC for 100 V power supply) for resistive load
Maximumpower:60W,480VA
Maximumvoltage:30VDC,264VAC(132VACfor100Vpowersupply)
Maximumcurrent:2ADC/AC
Contact input: Voltage-free contact, 1 point
:RemoteswitchingforsamplegassuctionpumpON/OFF
Self-diagnostics
Error (failure) : Sensor unit error, heater temperature error, temperature sensor disconnection,
device temperature error, CPU error, fan stop.
Alarm (warning) : Heater unstable, sensor defect, electromotive force abnormal, asymmetry voltage
error, calibration error, sensor resistance error, O
2
concentrationupper/lower,overrange.
Serial communication
Comm. signal : RS-232 , one way
Baud rate : 38,400 bps
Data (ASCII) : O
2
concentration,unit,alarm/error
Calibrationmethods:
(1) 3 point: 10 ppm, 1000 ppm, Air
(2) 2 point: zero and span calibration may be set freely
(3) 1 point:
(4) Air calibration
Warm-up time : Within 20 min
Power supply:
Powersupply :100-120VAC/200-240VAC,50/60Hz
Acceptable range: 100 to 120 V AC ±10%,50/60Hz
200to240VAC±10%,50/60Hz
Power consumption
:100to120VAC,200VAmax.
200to240VAC,400VAmax.
Dimensions :213(W)x132(H)xapprox.375(D)mm
Weight:Approx.5kg
Finish: Polyester coating