<2.Specications>
2-1
IM 11M10B01-01E
2. Specications
2.1 StandardSpecications
Measurement object
:Oxygenconcentrationsininertgasescontainingnoammablegas,silica,corrosivegas,orliquid
(including water vapor).
Measurement system: Zirconia system
Samplingmethod: Pump, aspirator, or no suction device.
Pumpandaspiratorsuctionowrate
:Approx.1.0l/min.
Aspirator suction conditions
: Air or N
2
,supplypressure65to100kPaG,totaldischargeow10l/minmax.(whengasinletand
outlet are at ambient atmospheric pressure).
Sample gas conditions
Flowrate :200±25ml/min(onlyappliestosensor).
Temperature : 0-50°C (non-condensing).
Humidity : Non-condensing.
Pressure : 0-300 PaG
Measurement range: 0-10 ppm O
2
to 0-100 vol% O
2
Resolution: 0.01 ppm O
2
.
Display : 4 digit LED.
Main display: O
2
concentration (auto switching).
Subdisplay:Parameteroralarm/errornumber
Unit : %, ppm.
Output range
Auto : 0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-1%, 0-10%, 0-100% (default)
Other : 0-0 ppm, 0-00 ppm, 0-000 ppm, 0-%, 0-0% can be set,
is an integer from 2 to 9.
Fixed :Setto0-10ppm,0-100ppm,0-1000ppm,0-1%,0-10%,or0-100%.
Partial : Lower value or upper value of range can be set.
Note:Span(uppervalue-lowervalue)is20%FSormoreofabovexedrange.
Example:200-400ppmwhenxedrangeis0-1000ppm,20-40ppmwhenxedrangeis0-
100 ppm.
Analog output: 2 outputs
Primary :4to20mADC(maximumloadresistance:550Ω)
Secondary :Selectfrom0-1,0-5,0-10VDC(loadresistance:10kΩorgreater)