Owner's manual
3
IM 77J08A04-01E 1st Edition  Nov, 18. 2011-00
3.  EXTERNAL WIRING
  WARNING
Be sure to turn OFF the power supply before wiring to 
avoid the risk of electric shock. Use a tester or similar 
device to ensure that no power is being supplied to a 
cable to be connected.
M4 screw terminals are provided for the connection of external 
signals. Attach a crimp-on lug to each wire for connection to the 
terminals.
 • Recommended cables: A nominal cross-sectional area of 
0.5 mm
2
 or thicker for signal cables, and that of 1.25 mm
2
 or 
thicker for power cables.
Wiring Diagram
1
2
4 5
6 7
6
7
4
5
+
−
Output
Input
Power Supply
+
–−
1
2
PS+
−
+
−
  NOTE
 ● Do not connect anything to the terminals that are 
not used in the wiring diagram. Otherwise it may 
cause the malfunction or damage.
 ● The power line and input/output signal lines should 
be installed away from noise-generating sources. 
Otherwise accuracy cannot be guaranteed.
 ● Adhere strictly to the specications to avoid 
overheating or damage. Before turning on the 
power, ensure the following:
•• Power supply voltage and input signal value 
applied to the product should meet the required 
specications.
•• The external wiring to the terminals and wiring to 
ground are as specications.
 ● Do not operate the product in the presence of 
ammable or explosive gases or vapors.
 ● This product is sensitive to static electricity; 
exercise care in handling. Before you operate the 
product, touch a nearby metal part to discharge 
static electricity.
Power Supply and Isolation
Power supply voltage: 24 V DC±10% (percentage ripple: less 
than 5% p-p)
Current consumption : 65 mA
Insulation resistance: 100 MΩ at 500 V DC between (input and 
output) and power supply.
Withstand voltage: 500 V AC/min. between (input and output) 
and power supply
4.  MAINTENANCE
The product starts running immediately when the power is turned 
on; however, it needs 10 to 15 minutes of warm-up before it 
meets the specied performance.
4.1  Calibration Apparatus
 • A voltage current source (Yokogawa GS200 or equivalent)
 • A digital multimeter, DMM (Yokogawa 7561 or equivalent)
 • A precision resistor of 250 Ω±0.01%, 1W
4.2  Calibration Procedure
1. Connect the instruments as shown below. 
1
2
4 5
6 7
2
4
5
+
–
+
–
Ro
DMM
L+
N–
Power 
supply
24V DC
Voltage
current
source
Ro: a precision resistor 250Ω
Input
Output
Ro is used for current output (FA4A).
2. Use the voltage current source and apply the electromotive 
force equivalent to 0, 25, 50, 75, and 100% of the measuring 
range to the distributor. 
3. Verify that the corresponding output voltages are 0, 25, 50, 
75, and 100% respectively and within the specied accuracy 
rating.