6.2 Calibration setting
This mode is used to set calibration concentration and actions. The calibration setting involves
calibration concentration, zero calibration, calibration range and auto calibration component.
6.2.1 Setting of calibration concentration
It allows you to set concentrations of the standard gas (zero and span) of each channel used for
calibration.
MODE
( )
ENT
( )
ENT
( )
ENT
(1) During measurement, press the
MODE
key to
display the User mode.
(2) Point the cursor to “Calibration
Parameters” by pressing the
or key.
Press the
ENT
key.
(3) In the “Calibration Parameters” screen
that appears, point the cursor to “Calibra-
tion Value” by pressing the or
key. Press the
ENT
key.
(4) In the “Calibration Concentration Ch
Selection” screen that appears, point the
cursor to Ch you want to set by using the
or key. Press the
ENT
key.