User's Manual IR200 Infrared Gas Analyzer IM 11G02M01-01E IM 11G02M01-01E 3rd Edition
PREFACE We are grateful for your purchase of Yokogawa’s Infrared Gas Analyzer (Model : IR200). • First read this instruction manual carefully until an adequate understanding is acquired, and then proceed to installation, operation and maintenance of the analyzer. Wrong handling may cause an accident or injury. • The specifications of this analyzer are subject to change without prior notice for further product improvement.
CAUTION ON SAFETY First of all, read this “Caution on safety” carefully, and then use the analyzer in the correct way. • The cautionary descriptions listed here contain important information about safety, so they should always be observed. Those safety precautions are ranked in 3 levels, “DANGER”, “CAUTION” and “PROHIBITION”. DANGER Wrong handling may cause a dangerous situation, in which there is a risk of death or heavy injury.
Caution on piping DANGER In piping, the following precautions should be observed. Wrong piping may cause gas leakage. If the leaking gas contains a toxic component, there is a risk of serious accident being induced. Also, if combustible gas is contained, there is a danger of explosion, fire or the like occurring. • Connect pipes correctly referring to the instruction manual. • Exhaust should be led outdoors so that it will not remain in the locker and installation room.
Caution on use PROHIBITION • Do not allow metal, finger or others to touch the input/output terminals in the instrument. Otherwise, shock hazard or injury may occur. • Do not smoke nor use a flame near the gas analyzer. Otherwise, a fire may be caused. • Do not allow water to go into the gas analyzer. Otherwise, hazard shock or fire in the instrument may be caused.
r After - Sales Warranty d Do not modify the product. d During the warranty period, for repair under warranty carry or send the product to the local sales representative or service office. Yokogawa will replace or repair any damaged parts and return the product to you. d Before returning a product for repair under warranty, provide us with the model name and serial number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.
CONTENTS PREFACE.............................................................................................................................. i CAUTION ON SAFETY .................................................................................................... ii r After - Sales Warranty ................................................................................................... v 1. OVERVIEW ..........................................................................................................
7. MAINTENANCE ......................................................................................................... 7-1 7.1 7.2 7.3 7.4 7.5 Daily check ...................................................................................................... Daily check and maintenance procedures ...................................................... Replacement of power fuse............................................................................. Moisture interference compensation adjustment ........
1. OVERVIEW The infrared gas analyzer (IR200) measures the concentrations of CO2, CO, CH4, SO2, NO and O2 contained in sample gas. CO2, CO, CH4, SO2, and NO are measured by non-dispersion infrared method, while O2 is measured by paramagnetic or fuel cell method. Up to 4 components including O2 (3 at the maximum for measurement of components other than O2) can be measured simultaneously. A high-sensitivity mass flow sensor is used in the detector unit of infrared method.
2. NAME AND DESCRIPTION OF EACH PART 2.
3. INSTALLATION DANGER This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent explosion, fire or other serious accidents. CAUTION • For installation, observe the rule on it given in the instruction manual and select a place where the weight of gas analyzer can be endured. Installation at an unsuited place may cause turnover or fall and there is a risk of injury. • For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an injury.
3.2 Installation of analyzer There are two methods of installing the analyzer. For detailed dimensions, see Chapter 9.3. (Unit:mm) 19 inch rack mounting (Note) Types External dimensions Mounting dimensions Mounting method Guide rail MODE ENT Slide rail M6 ZERO ESC SPAN 429 450 or more 483 465 101.
Observe the following when connecting the gas pipes. • The pipes should be connected to the gas inlet and outlet at the rear panel of the analyzer, respectively. • Connect the sampling system to the instrument by using corrosion-resistant tube such as Teflon, stainless steel, or polyethylene. In case where there is no danger of corrosion, don't use rubber or soft vinyl tube. Analyzer indication may become inaccurate due to the adsorption of gases.
(2) Sampling gas flow rate A flow rate of sampling gas must be 1 ± 0.5L/min. A flow meter should be provided as shown in Fig. 3-2 to measure flow rate values. (3) Preparation for standard gas Prepare the standard gas for zero/span calibration.
(7) Example of sampling system configuration The system configuration may vary depending upon the nature of measured gas, coexistent gases or application. A typical configuration diagram is shown in Fig. 3-2. Since a system configuration depends upon measured gas, consult with Yokogawa.
3.5 Wiring method Caution on wiring • Wiring work must be performed with the main power set to OFF to prevent electric shocks. CAUTION • Enforce construction of class-D grounding wire by all means, 100V or less. If the specified grounding construction is neglected, a shock hazard or fault may be caused. • Wires should be the proper one meeting the ratings of this instrument. If using a wire which cannot endure the ratings, a fire may occur. • Be sure to use a power supply of correct rating.
When noise source is in the vicinity Do not install the instrument near an electric device that generates noise from the power supply (such as high-frequency furnace and electric welder). If there is no other choice but to install it near such devices, provide completely separate power line to eliminate noise. In case noise may enter from a relay, solenoid valve, etc. through power supply, connect a varistor or spark killer to the noise source as shown below.
(4) Contact input (DI): terminal block 2, 13 to 20 , terminal block 3, 5 to 10 . • It is for a contact input at no voltage. An input is provided when switching to short circuit (on) or open (off). • No voltage is applied to the terminals. (5) Contact output (DO): terminal block 3, 13 to 20 , terminal block 4 and terminal block 5 • Contact rating: 250 V AC/2 A, load resistance • An output is for a relay contact output. An output is provided when switching to conductive (on) or open (off).
(7) List of terminal blocks 1 to 5 Terminal block 1 CH5 output 2 (AO) 1 1 11 2 12 CH4 output 2 (AO) 1 3 13 4 14 CH3 output 2 (AO) 1 5 15 6 16 CH2 output 2 (AO 1 7 17 8 18 CH1 output 2 (AO) 1 9 19 10 20 Terminal block 2 Unused Unused *1 *1 2 CH8 output 1 (AO) *2 2 O2 sensor input 1 *1 Unused *1 Unused *1 2 CH7 output 1 (AO) Unused 2 CH6 output 1 (AO) Unused *1 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 Unused CH4 remote range c
(8) Description on terminal block Terminal block 1 Terminal block for analog output (non-isolated output) Terminal block 1 CH5 output 2 (AO) 1 1 2 12 CH4 output 2 (AO) 1 3 13 4 14 CH3 output 2 (AO) 1 5 15 6 16 CH2 output 2 (AO 1 7 17 8 18 9 19 10 20 CH1 output 2 (AO) 1 11 Unused Unused Output : 4 to 20 mA or 0 to 1 V DC Between 1 – 2 : CH5 output Between 3 – 4 : CH4 output Between 5 – 6 : CH3 output Between 7 – 8 : CH2 output 2 CH8 output 1 (AO) Between 9 – 10
Terminal block 3 Terminal block 3 1 11 2 12 3 13 4 14 Remote hold input (DI) 5 15 6 16 Average value reset input (DI) 7 17 8 18 9 19 10 20 Unused Unused Auto calibration remote start input (DI) (M3.5 screw) Between 1 to 4 : For internal connection. Must not be wired. (Must not be used as junction terminal.) Between 5 – 6 : Remote hold input. No hold when open. Output hold when short-circuited. Between 7 – 8 : Average value reset input.
Terminal block 4 Terminal block 4 Contact output for CH4 span calibration (DO) 1 2 12 Contact output for CH3 span calibration (DO) 3 13 4 14 Contact output for CH2 span calibration (DO) 5 15 6 16 Contact output for CH1 span calibration (DO) 7 17 8 18 Contact output for zero calibration (DO) 9 19 10 20 11 (M3.
Terminal block 5 CH3 alarm output (DO) CH2 alarm output (DO) CH1 alarm output (DO) 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 : For internal connection. Must not be wired. (Must not be used as junction terminal.) Between 2– 3 : CH3 alarm output. Conductive at 2– 3 and open at 3– 4 when set value is exceeded. Open at 2– 3 and conductive at 3– 4 otherwise. and 3– 4 Terminal block 5 Unused 1 and 11 – 14 Between 5– 6 Unused 9 19 10 20 (M3.
(9) Timing of calibration contact output 1. Manual calibration (See “6.9 Calibration”). ZERO ENT ENT • Zero calibration Calibration end Pump ON/OFF contact on off on Zero calibration contact off CH1 to CH4 span calibration contact off Output hold function Output signal hold (with hold ON setting) Hold extension time When selecting CH2 using and keys.
4. OPERATION 4.1 Preparation for operation (1) Check of gas sampling tube, exhaust tube and wiring Check that the pipes are correctly connected to the gas sampling port and drain port. Check that the analyzer is correctly wired as specified. 4.2 Warm-up operation and regular operation (1) Operation procedure 1. Turn ON the power switch at the left of the front panel. In one or two seconds, the measurement screen will appear at the front panel. 2.
5. DESCRIPTION OF DISPLAY AND OPERATION PANEL This section describes the display and operation panel of the gas analyzer. It also explains the name and description of function on the operation panel. See Fig. 5-1. 5.1 Name and description of operation panel Display INFRARED GAS ANALYZER POWER Power switch MODE ZERO ESC ENT SPAN Controls • Display : The measurement screen and the setting items are displayed. • Controls : The configuration is as shown below.
5.2 Overview of display and operation panel • Measurement mode *1 *1) The screen configuration depends upon the display channel. When channel 5 or later is displayed on the screen, scroll the screen by using the UP or DOWN key. ZERO ZERO Calibration • Measurement mode SPAN Calibration SPAN MODE ESC • User mode ESC ESC • User mode 1. Changeover of Range 2. Setting about Calibration 3. Alarm Setting 4. Setting of Auto Calibration 5. Setting of Auto zero calibration 6. Setting of Peak Alarm 7.
5.3 Overview of display screen (1) Measurement mode screen On turning on the power switch, the Measurement Mode screen will appear. The measurement screen depends on the number of components. The following screen configuration shown as an example is for CO2, CO, O2 (Output at channel 6). When channel 5 or later is displayed, scroll or key to view another configuration which is beyond the screen.
• O2 correction concentration average value: AV CH (component) where “ CV ** ” is displayed as “ AV CV CO” in the component display and O2 average value, a value obtained by averaging O2 conversion concentration value or O2 concentration value in a fixed time is output every 30 seconds. Averaging time can be changed between 1 minute and 59 minutes or 1 hour and 4 hours according to the average time settings (See 6.7 Parameter setting).
(3) Contents of measured channel (CH) The contents in each measured CH corresponding to the type are given below: Suffix/Option Code Measurable O2 O2 correction component analyzer Output and Corresponding Channel CH1 CH2 -A -B -C -D -E -F -G -H -J -K -L -M -A -B -C -D -E -F -G -H -J -K -L -M -A -B -E -F -G -J -K N N N N N N N N N N N N 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 Not specified Not
5.4 General operation • Measurement mode The measurement mode can be displayed for up to 5 channels in a single screen. When viewing a channel beyond the 5 channels, or ) key and the press the ( screen can be scrolled one by one channel at a time. ZERO See 6.9.1, Zero calibration See 6.9.2, Span calibration SPAN MODE • ESC User mode displays the following settings.
6. SETTING AND CALIBRATION 6.1 Changeover of range This mode is used to select the ranges of measured components. 1. During measurement, press the MODE MODE key to display the User mode. 2. Point the cursor to “Changeover of Range”. Press the ENT key. ENT 3. The “Channel Selection” screen appears. “ or “ ” key until the “ ” Press the “ cursor moved selects a desired CH (component). 4. After selection, press the ENT key.
6.2 Calibration setting This mode is used to set calibration concentration and actions. The calibration setting involves calibration concentration, calibration gas, zero calibration, calibration range and auto-calibration component. “Calibration Gas” is displayed if NO or SO2 is included in components to be analyzed. 6.2.1 Setting of calibration concentration It allows you to set concentrations of the standard gas (zero and span) of each channel used for calibration. MODE 1.
5. In the “Calibration Concentration Selection” screen that appears, select any concentration item you want to set by pressing the , key. Note) Analyzers other than the zirconia O2 instrument cannot perform zero setting. Cursor for setting value ENT 6. In the “Calibration Concentration Value Setting” screen that appears, enter calibration gas concentration values (zero and or span). For value entry, press the key, and a 1-digit value increases or decreases. By pressing the , the digit moves.
6.2.2 Setting of calibration gas Whether moisture is contained in zero gas and span gas used for calibration can be set in advance as follows. Make the setting according to the sampling system created. If sampling system controls the moisture content in the calibration gas (ZERO or SPAN) to become the same as that in the sample gas to be fed to the analyzer using an electronic cooler, etc., set “With” water vapor. MODE (1) Press the MODE key in measurement state to display the User mode.
(5) On the “Gas Selection” screen that appears, select “with” or “without” of moisture content in calibration gas using or the key. the • Select “with” to have the calibration gas contain a certain amount of moisture. • Select “without” to use dry gas (moisture content: saturated at 70°C or lower) such as cylinder gas as calibration gas. • When “without” is selected for zero gas, “with” cannot be selected for span gas. After the selection is made, press the key.
6.2.3 Setting of manual zero calibration If zero calibration is to be made manually, select whether to calibrate all components at once or each of them separately upon selection. MODE 1. During measurement, press the to display the User mode. MODE key 2. Point the cursor to “Setting about Calibration” by pressing the or key. key. Press the ENT ( ) ENT ( ) ENT ( ) 3. In the “Setting about Calibration” screen that appears, point the cursor to “About or ZERO Calibration” by pressing the key.
5. In the “Zero Calibration Selection” screen that appears, select “at once” or or key. “each” by pressing the When selecting “at once”, the CH (components) to be set can be zero-calibrated at the same time. When selecting “each”, either of the CH (components) to be selected is zero-calibrated. After setting, press the key, and the calibration you specified is carried out. ENT To close “About ZERO Calibration” To close “About ZERO Calibration” setting or to cancel this mode midway, press the ESC key.
6.2.4 Setting of calibration range This mode is used to set if the range of each CH (component) at the zero and span calibration (manual calibration or auto calibration) should be calibrated with a single range or 2 ranges. MODE 1. During measurement, press the to display the User mode. MODE key 2. Point the cursor to “Setting about Calior key. bration” by pressing the Press the key. ENT ( ) ENT ( ) ENT 3.
5. In the “Calibration Range Selection” screen that appears, select “both” or “current” by or key. pressing the • When selecting “both”, Range 1 and Range 2 of the set CH are calibrated together. • When selecting “current”, the range alone displayed at the set CH is calibrated. To close Setting of Calibration Range To close Setting of Calibration Range or ESC key. to cancel this mode midway, press the ESC A previous screen will return.
6.2.5 Setting of auto-calibration component It sets the CH (component) to be calibrated in the auto-calibration. MODE 1. During measurement, press the to display the User mode. MODE key 2. Point the cursor to “Setting about Calior key. bration” by pressing the key. Press the ENT ( ) ENT ( ) ENT 3. In the “Setting about Calibration” screen that appears, point the cursor to “Auto Calibration Components” by pressing the or key. Press the key. ENT 4.
5. In the “Auto Calibration component Selection” screen that appears, select “enable” or “disable” by pressing the or key. After setting, press the key. ENT To close Auto Calibration Component setting To close “Setting of Auto Calibration Component” or to cancel this mode midway, press the ESC ESC key. A previous screen will return. ( ) ENT End of Auto Calibtation Component Setting Example Auto calibration is made in the following rules according to the setting. 1.
6.3 Alarm setting 6.3.1 Setting of alarm values This mode is used to set the upper and lower limit value to provide an alarm output against measured concentration during measurement. Before changing the alarm setting, set the ON/OFF to OFF. MODE 1. During measurement, press the to display the User mode. MODE key 2. Point the cursor to “Alarm Setting” by pressing the or key. Press the key. ENT ( ) ENT 3.
After setting, the alarm setting is now key. completed by pressing the Cursor for setting value ENT To close the Alarm Setting To close the " Alarm Setting" or to cancel ESC key. this mode midway, press the ESC A previous screen will return. Setting Range 0 to 100%FS (Settable in each range) ENT End of Alarm Setting Description of setting items Upper limit value: Sets the upper limit value (concentration) of alarm. Lower limit value: Sets the lower limit value (concentration) of alarm.
6.3.2 Hysteresis setting To prevent chattering of an alarm output near the alarm setting values, set hysteresis. 1. In the “Alarm Setting CH Selection” screen that appears, point the cursor to or “Hysteresis” by pressing the key. Press the key. ENT ( ) ENT 2. In the “Hysteresis Value Setting” screen that appears, enter hysteresis values. For the value entry, 1-digit value is increased or decreased by pressing the or key, and pressing the key moves the digit.
6.4 Setting of auto calibration 6.4.1 Auto calibration Auto calibration is automatically carried out when zero calibration and span calibration are set. MODE Before changing the setting of auto calibration, set the ON/OFF to “OFF”. 1. During measurement, press the to display the User mode. MODE key 2. Point the cursor to “Setting of Auto or Calibration” by pressing the key. Press the key. ENT 3.
The auto calibration status contact output is closed while auto calibration is in progress (hold extension time by the length of time the same as gas replacement time included if the hold setting is set to “ON”), and open in other cases. Example Start Time Cycle Flow Time ON/OFF SUN 1 350 12:00 day sec ON In case where auto calibration is carried out at the above setting.
6.4.2 Forced stop of auto calibration This mode is used to cancel the auto calibration forcedly. 1. In the User mode that is displayed, point the cursor to “Setting of Auto Calibraor key. tion” by pressing the Press the key. ENT ( ) ENT 2. In the “Setting of Auto Calibration” item selection screen that appears, point the cursor to “Stop Auto Calibration” by pressing the or key. Press the key. ENT ( ) ENT 3. “Stop Auto Calibration” is inverted.
“Auto Calibration” screen Example In case where setting the auto calibration components (see Item 6.2.5) to “CH1: enable” and “CH2: enable”. • Zero calibration A message, “Zero calibration” blinks at CH1 and CH2. • CH1 span calibration A message, “Span calibration” blinks at CH1. • CH2 span calibration A message, “Span calibration” blinks at CH2. Caution During auto calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Forced Stop of Auto Calibration”.
6.5 Setting of auto zero calibration 6.5.1 Auto zero calibration Auto zero calibration is automatically carried out when zero calibration are set. Components for which a calibration is to be made are determined by setting of auto calibration component in 6.2.5. MODE Before changing the setting of auto zero calibration, set the ON/OFF to “OFF”. 1. During measurement, press the to display the User mode. MODE key 2. Point the cursor to “Setting of Auto Zero or Calibration” by pressing the key.
The auto calibration status contact output is closed while auto zero calibration is in progress (hold extension time by the length of time the same as gas replacement time included if the hold setting is set to ON), and open in other cases. Example Start Time SUN 12:00 Cycle Flow Time 12 300 hour sec ON/OFF ON In case where auto zero calibration is carried out at the above setting.
6.5.2 Forced stop of auto zero calibration This mode is used to cancel the auto zero calibration forcedly. 1. In the User mode that is displayed, point the cursor to “Setting of Auto Zero or Calibration” by pressing the key. Press the key. ENT ( ) ENT 2. In the “Setting of Auto Zero Calibration” item selection screen that appears, point the cursor to “Stop Auto Zero Calibration” by pressing the or key. Press the key. ( ) ENT 3. “Stop Auto Zero Calibration” is inverted.
“Auto Zero Calibration” screen Example In case where setting the auto calibration components (see Item 6.2.5) to “CH1: enable” and “CH2: enable”. • Zero calibration A message, “Zero calibration” blinks at CH1 and CH2. Caution During auto zero calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Forced Stop of Auto Zero Calibration”. When the key lock is set at “ON”, even the “Forced Stop of Auto Zero Calibration” cannot be used.
6.6 Peak alarm setting When the number of peaks of which CO concentration exceeds the upper limit value exceeds the setting time, a peak alarm is outputted. This section describes how to perform various settings of peak alarm. Note) The setting is optional and is valid only when peak alarm function is provided. MODE 1. Press the key in the Measurement mode, and the User mode appears. MODE 2. Point the cursor to “Setting of Peak Alarm” by pressing the or key. Press the key.
Setting range • Alarm Value • Alarm Count • Hysteresis : 0 to 1000 ppm 5 ppm step (initial value: 500 ppm) : 1 to 99 times (initial value: 5 times) : 0 to 20% of full scale (initial value: 0% of full scale) * The hysteresis setting is made in terms of full scale. Action of peak alarm Example (ON) (OFF) If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over the set times per hour, a peak count alarm becomes closed (ON).
6.7 Parameter setting It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be set are as follows: Description of setting items • Current Time • Key Lock • Remote Range • Output Hold • Reset Average Output • Response Time • Average Period • Maintenance mode : Sets the current day of the week, hour and time. Note: The clock backup time is 2 days. If power is turned on after it is kept off for 2 days or louger, make the time setting again.
4. In the Parameter Setting screen that appears, enter the numeric values and set the items. Entering the numeric values or setting the items should be carried out by using the or key. To move the key. cursor to the right, press the key, that the After setting, press the parameter setting is carried out with the value you set. ENT To close Parameter Setting To close the "Parameter Setting" screen or cancel this mode midway, press the ESC key. A previous screen will return.
(2) Auto calibration Auto calibration start Auto calibration end Calibration Hold extending time Output hold Time hold set to the gas flow time (3) External hold Strapped Remote hold Open Output hold (4) Screen display during holding The “Holding” message blinks on the measuring screen. Since the screen displays the process of calibration is displayed during the manual calibration, “Holding” is not displayed even if it is holding, but the screen is displayed with the hold extending time.
Average period It allows you to set an average period of the average value of O2 correction. It enables you to set an average time of 1 to 59 minutes (1-minute step) or 1 to 4 hours (1-hour step). Changing the setting resets the average value of O2 correction and O2 average value. (Pressing the ENT validates the resetting only for components whose setting was changed.) Example of average action Suppose the average period is 1 hour. Reset Average value Time Time • Sampling occurs every 30 seconds.
6.8 Maintenance mode Check of sensor input value, display of error log file, setting of O2 correction reference value, adjustment of moisture interference, setting of auto off time of display backlight, and password setting can be performed as follows. A password is selected first. The password must be entered to perform measurement from next time onward. Select the maintenance mode in “6.7 Parameter setting,” and this screen is displayed.
• O 2 correction reference value setting screen Description of O2 corection reference value setting screen O2 correction value is calculated from the following equation by values set herein: O2 correction concentration value = 21 set value 21 - O2 concentration measured value component x concentration measured value The value of the fractional part in this calculation is limited to 4. If it is 4 or more, it is fixed at 4.
• Moisture interference compensation adjustment screen Description of moisture interference compensation adjustment screen By setting the effect of moisture content in measured gas on this screen, the value of the effect is subtracted to find the measurement value. Adjustment is carried out by one of the following methods. • Select a desired channel (CH) and change the • value displayed at right. Select ALL and perform auto calculation for all the channels in batch.
6.9 Calibration ZERO 6.9.1 Zero calibration It is used for zero point adjustment. For zero calibration gas, see 3.4 (3), Preparation for standard gas in Sampling. Use a gas according to application. 1. Press the ZERO key on the Measurement screen to display the Manual Zero Calibration screen. 2. Select the CH (component) to be calibrated by pressing the or key. After selection, press the key, and zero gas will be supplied.
6.9.2 Span calibration It is used to perform a span point adjustment. Supply calibration gas with concentration set to the span value to perform the span calibration. For the span calibration gas for the NO, SO2, CO2, CO, CH4 masurement, use the standard gas with a concentration of 90% or more of the range value. SPAN For the calibration gas for the O2 measurement, use the standard gas of 1 to 2 vol%. 1. Press the SPAN key on the Measurement screen to display the Manual Span Calibration screen. or 2.
7. MAINTENANCE 7.1 Daily check (1) Zero calibration and span calibration 1. It is used for zero point adjustment. For calibration, refer to 6.9.1, Zero calibration. 2. After zero calibration, perform span calibration. For calibration, refer to 6.9.2, Span calibration. 3. Zero calibration and span calibration should be performed once a week, if required. (2) Flow check 1. Sampling gas flow rate and purge gas flow rate should be as follows; Sampling gas flow rate: 1.0±0.
7.3 Replacement of power fuse Note) Before fuse replacement, locate the probable cause of a blown fuse. 1. Turn OFF the power switch. 2. Detach two mounting screws fastening to the top cover. Mounting screws (2 places) INFR ARED GAS ANAL YZER POWER MOD E ESC ZER O ENT SPAN Power switch 3. Slide the top cover backward, and pull it upward. Top cover INFR ARED GAS ANALYZE R POWER MOD E ESC ZER O ENT SPAN 4. 5.
7.4 Moisture interference compensation adjustment (1) Be sure to carry out moisture interference compensation adjustment at the following timing when NO or SO2 is included in components to be analyzed. 1) Before starting operation for the first time after the analyzer is incorporated into the system 2) At periodic inspection to be performed at 6 month to 1 year intervals 3) After the cell is cleaned (2) Adjustment procedure (See “Moisture Interference Compensation Adjustment screen” in “6.
7.5 Cleaning of measuring cell Entry of dust or water drops in the measuring cell contaminates the interior of the cell, thus resulting in a drift. Clean the inside if dirty. Then, check the sampling device, especially the filter, to prevent the cell from being contaminated by dust or mist. 7.5.1 Disassembly and assembly of measuring cell There are two kinds of measuring cells, on block cells (cell length: 4 mm, 8 mm, 16 mm, 32 mm) and pipe cells (Cell length: 64 mm, 125 mm and 250 mm).
No.
(2) How to remove block cell (See Fig. 7-2) 1. For Step 1 to 4., see Item 7.5.1, (1) How to remove pipe cell. 5. Remove the connector to the detector output cord from the printed board. For the 2-component analyzer, remove the output cord connector of the 2-component analyzer detector (No. 13) from the printed board, and then remove the 2-component detector by unscrewing two mounting screws (No. 14) fastening the 2-component detector. 6. Unscrew the two screws (No.
No. 1 (2) Name Screw (for fixing light source unit) Filter 3 Screw (for fixing detector) 4 Base plate 5 Light source unit 6 Screw (for fixing block cell) 7 Block cell 8 Infrared transmission window (window holder) 9 O-ring 10 Screw (for fixing base plate) 11 Chopper motors connector 12 Detector (13) Detector: Install in the 2-component analyzer (14) Screws: For mounting the 2-component detector (14) (13) 10 6 12 1 (2) 8 9 7 3 5 11 4 Fig.
3) How to remove measuring unit (See Fig. 7-3) 1. For Step 1. to 4., see Item 7.5.1(1), How to remove pipe cell.5. Remove the detector output cord connector from the printed board. 6. Remove wiring to the 2-pin terminals of the infrared ray light source assembly and chopper motor pin connector (No. 8) from the printed board. 7. Detach the 4 screws (No. 3) fastening the base plate (No. 4) to remove the measuring unit.
7.5.2 How to clean cell 1. To clean the cell inside or infrared ray transmission window, first clear large dirt of it with a soft brush and then wipe with soft cloth lightly. Don’t use hard cloth. Note) Handle the fragile window with care. Use care not to rub off the dirt from the window roughly. 2. If the window or the cell interior is very dirty, use a soft cloth moistened with absolute alcohol. 3.
d Precautions to be taken while checking 1. When handling reference gas (during calibration), carefully read the reference-gas instruction manual to use the gas correcrtly. In particular, special attention must be taken in handling carbon monoxide gases; otherwise, you may suffer from gas poison-ing. 2. During maintenance checks, be sure to keep the near fan on. If any gas leaks, you may suffer from gas poisoning. 3.
8. TROUBLESHOOTING 8.1 Error message If errors occur, the following contents are displayed. Error display Error contents Probable causes Error No.1 Sector motor rotation detector signal is faulty. • Infrared ray light source is faulty. • Sector motor rotation is faulty or stopped. • Motor rotation detector circuit is faulty. • Amplifier circuit is faulty. Error No.3 A/D conversion signal is faulty. • Circuit is faulty. Error No.4 Zero calibration is not within the allowable range.
In case of Error No. 5 and No. 7 ENT ? Pressing ESC delete the error display. ENT ESC Calibration is continued. Unless another calibration error occurs, calibration is carried out to the end, the Measurement screen returns. Error log file If error occurs, the history is saved in an error log file. The error log file exists in the maintenance mode. Error log screen Day of the week and time when an error occurred. Component with which the error occurred.
8.2 Other Troubleshooting The following table shows how to remedy other troubles, such as faulty readings. Symptom Checking Item Remedy, etc. Drift * Check if the sample gas is supplied to the analyzer at the specified flow rate. * Check the optical system, e.g., the sample cell window, O-ring, detector window and inside of the cell for heavy dirt.
9. SPECIFICATIONS 9.
(2) Standard functions Output signal holding: Output signals are held during manual and auto calibrations by activation of holding (turning on its setting). The values to be held are the ones just before start calibration mode. Indication values will not be held. Remote output holding: Output signal is held at the latest value by short-circuiting the remote output holding input terminals. Holding is maintained while the terminals are short-circuited. Indication values will not be held.
(3) Optional functions O2 correction: Conversion of measured CO , SO2 and NO gas concentrations into values at reference O2 concentration. 21-On Correction formula: C = –––––––– x Cs 21-Os C : Sample gas concentration after O 2 correction Cs : Measured concentration of sample gas Os : Measured O2 concentration O n : Standard O 2 concentration (value changeable by setting) *The upper limit value of the fractional part in this calculation is 4.
(6) Installation Requirements • Use this instrument .(Select a place where the equipment does not receive direct sunlight, draft/rain or radiation from hot substances.) • Avoid a place where receives heavy vibration. • Select a place where atmospheric air is clean.
9.
Model and Suffix Codes Model Suffix code IR200 2nd Component 2nd Range (note 2) 3rd Component 1st Range (note 2) C D E F G H K L M P R T N 021000 ppm 022000 ppm 022500 ppm 025000 ppm 021% 022% 025% 0210% 0220% 0225% 0250% 02100% Not available 02500 ppm 021000 ppm 022000 ppm 022500 ppm 025000 ppm 021% 022% 023% 025% 0210% 0220% 0225% 0240% 0250% 0270% 02100% Not available 021000 ppm 022000 ppm 022500 ppm 025000 ppm 021% 022% 025% 0210% 0220% 0225% 0250% 02100% Not available C D E F G H K L M P R T N 0
3: Only available for CO measurement. 4: Only available for CO2, CO, SO2 or NO measurements. 5: O2 correction is available only for CO, SO2, and NO. Both average value output after O2 correction and average O2 value output are provided at the same time. A peak count alarm can be provided only for CO measurement. 6: Suffix Codes “E” and “U” are power cables with different voltage rating and plug type. Select appropriate code according to the operating power supply voltage to be used in the field.
Table 4: Two-component analyzer (CO2 and CH4) 1st component (CO2),1st range 2nd component (CH4), 1st range C D E F CH4 H G 0 – 1000ppm 0 – 2000ppm 0 – 2500ppm 0 – 5000ppm 0 – 1% 0– 2% K L M P R T 0 – 5% 0 – 10% 0 – 20% 0 – 25% 0 – 50% 0 – 100% D 0 – 2000ppm ––– ––– ––– ––– ––– ––– ––– E 0 – 2500ppm ––– ––– ––– ––– ––– ––– F 0 to 5000ppm G 0 – 1% ––– ––– ––– ––– ––– ––– ––– ––– ––– ––– CO2 H 0 – 2% K 0 – 5% L 0 – 10% M 0 – 20% R 0 – 50% ––– T 0 – 100% ––– : Avail
Table 9.
9.3 External Dimensions (unit: mm) 365 355.6 123.8 111.1 38 (24) 101.5 254 336.5 500 max 88.5 M4 (for slide rail) Power Switch M5 MODE Applicable both for JIS 100 mm and EIA 101.6 mm 177 POWER 100.8 7.
(2) Accessory slide rail (unit: mm) Model : 305A-20/Accuride International Inc. Closed 530±3 Cabinet member 1.52t The same or less (22.7) Intermediate member Open (9.5) Drawer member 1.27t 15.9±0.5 12.7±0.3 (24.4) 35.3±0.3 25.4±0.3 508.0±0.8 476.2±0.3 463.5±0.3 352.8±0.3 111.1±0.3 12.7±0.3 4.535.3 4.535.3 123.8±0.3 238.1±0.3 365.1±0.3 15.9±0.5 19.1±0.3 505.0±0.
Dedicated Zirconia O2 Sensor (to be purchased separately) For O2 correction, the IR200 can accept linealized 0 to 1 V DC signal coming from an analyzer calibrated to 0 to 25% O2 of full scale. Dedicated zirconia O2 sensor, Model ZX8D, is available from Yokogawa. Measuring method: Zirconia system Measurable component and measuring range: Measurable component Minimum range Maximum range 025 vol% Oxygen (O2) 0225 vol% T12.EPS Repeatability: Within ±0.
NO2/NO Converter Power consumption: Approx. 85 VA Safety conforming standards: EN61010-1 Pollution degree 2 Installation category II Weight: Approx. 1.1 kg (K9350LE) Approx. 1.2 kg (K9350LF) Sample gas requirements: Dust/drain removed, gas temperature at 1508C or less Part number: K9350LE (Non-CE conformity) K9350LF (CE conformity) Mounting: Indoor surface mounting Target gases: General boiler exhaust gas, atmosphere Catalyst: Amount; 2 cm3 Replacement cycle; Approx. 12 months (at flow rate of 0.
Customer Maintenance Parts List Model IR200 Infrared Gas Analyzer Measuring cell (Next Page) 1 2 INFRAR ED GAS ANALYZ ER POWER MODE ESC ZERO ENT SPAN Item 1 2 Parts No. K9218RB K9358LM Qty. 1 1 Description Fuse (Time Lag Fuse) Fuse holder All Rights Reserved, Copyright © 2002, Yokogawa Electric Corporation. Subject to change without notice. CMPL 11G02M01-01E 1st Edition : July.
2 d Measuring Cell 5 4 5 4 8 7 8 11 11 3 12 10 10 2 9 9 Item Parts No. 1 K9358MG 2 K9358MH K9358MJ 3 K9358MW 4 K9358MX 5 K9358MY 6 K9358MZ 7 K9358NF 8 K9358NT 9 K9358NU 10 K9358NV 11 K9358NW 12 K9358PJ CMPL 11G02M01-01E Qty. 8 1 1 1 1 2 1 2 4 4 2 2 2 Description Screw Base plate Base plate for 250 mm O-ring for Block cell Screw Screw Cover Washer Screw Screw Support Pushing for pipe cell O-ring for pipe cell 1st Edition : July.
Revision Record Manual Title : IR200 Infrared Gas Analyzer Manual Number : IM 11G02M01-01E Edition Date Remark (s) 1s t Oct. 2002 Newly published 2nd Aug. 2006 Fully revised due to style change to S3 P. v., Added "After-Sales Warranty". Section 2.1, "Description of each unit": Changed Fig. 2-1. Section 3.4, (6), "Example of sampling system configuration": Changed Fig. 3-2 Section 5.1, "Name and description of operation panel": Changed Fig. 5-1. Section 5.
Edition Date Remark (s) Section 9.3, "External Dimensions," NO2/NO Converter: Changed "Part number" descriptions, added" Safety conforming standards" descriptions, deleted "CE Marking" descriptions, and changed "External Dimensions" drawings.