User Manual
IM 11M13A01-04E
8-5
8. Detailed Data Setting
F8.1E.EPS
Switch operations Display
Output hold time
during calibration
>  ENT 
b10
>  ENT 
CAL
>   ENT 
CAL
>   ENT 
SPAn Y
>   ENT 
21.00
%
>   ENT 
OPEn/20.84
>   ENT 
20.84
%
>   ENT 
ZEro Y
>   ENT 
0.98 %
>   ENT 
OPEn/0.89
>  ENT 
0.89 %
>   ENT 
CALEnd
>   ENT 
b10
>  ENT Measured-value display 
>
Figure 8.1 Definition During Calibration
In a semi-automatic calibration, “during calibration” is the time, starting when a calibra-
tion instruction is executed with an infrared switch or a contact input, to make a series
of calibrations, until the preset output stabilization time elapses.
In an automatic calibration, “during calibration” is the time, starting when automatic
calibration is carried out at the calibration start time, until the preset output stabilization
time elapses.
(4) “Error” appears when Error 1 to Error 4 are being issued