Table 7.14 Calibration Setup Procedure
T7.13E.EPS
Switch operation Display Description
Display after the password has been entered.
Set the zero-gas concentration. Switch the parameter code to B01.
Here, set 0.98%.
Touch the [ENT] key to display the currently set value.
Touch the [>] key to move the position of the digit that is flashing to 1.
Touch the [ ] key to change to 0.
Touch the [>] key to move the position of the digit that
is flashing to the right one digit.
Touch the [ ] key to change the numeric value to 9.
Touch the [>] key to move the position of the digit that
is flashing to the right one digit.
Touch the [ ] key to change the numeric value to 8.
Touch the [ENT] key to have all the digits flash.
Touch the [ENT] key again to stop the flashing.
Touch the [ENT] key once again to switch to the
parameter code selection display.
A01
b01
001.00
%
001.00
%
000.00
%
000.00
%
000.90
%
000.90
%
000.98
%
000.98
%
000.98
%
b01
The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.
b03
0
0
0
b03
Basic
panel
display
Next, set the calibration mode. Switch the parameter code to B03.
Touch the [ENT] key to display the currently set value.
If it is 0, you can leave it as is. If it is other than 0, change it to 0 (zero).
Touch the [ENT] key. The numeric value will flash.
Touch the [ENT] key again to stop the flashing.
Touch the [ENT] key once again to switch to the parameter code selection display.
Touch the [>] key together with the [ENT] key to return to the basic panel display.
(This is not required if you proceed to make another setting.)
(The displayed numeric characters indicate the measurement gas concentration.)
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
.
..
.
.
.
.
...
.
...
.
.
..
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
...
Set the span-gas concentration by above procedure, set 21 %.