User Manual
IM 11M13A01-04E
7-13
7. Startup
T7.6E.EPS
ENT 
A01
Display after the password has been entered.
ENT 
C01
Set the oxygen concentration at 4 mA. Change the parameter code to C11. 
Touch the [ ] key to switch to Group C.
ENT 
C01
Touch the [>] key to move the position of the digit that is flashing to the right.
ENT 
C11
Touch the [ ] key to enter the numeric value 1.
ENT 
000
Touch the [ENT] key to display the current set value (0% O2 has been set).
ENT 
000
Touch the [>] key to move the position of the digit that is flashing to the right.
ENT 
010
Touch the [ ] key to change the numeric value to 1. 
ENT 
010
If you touch the [ENT] key, all the digits flash.
ENT 
010
Touch the [ENT] key again to stop the flashing.
ENT 
C11
Touch the [ENT] key once again, and the display will switch to the parameter code selection
display.
ENT 
C11
Set the oxygen concentration at 20 mA. Touch the [>] key to move the position of the digit 
that is flashing to the right.
ENT 
C12
Touch the [ ] key to enter the numeric value 2.
ENT 
025
Touch the [ENT] key to display the current set value.
ENT 
025
Touch the [>] key to move the position of the digit that is flashing to the right.
ENT 
020
Touch the [ ] key to change the numeric value to 0. The numeric value will change from 
5 to 6 ... to 9 and back to 0.
ENT 
020
If you touch the [ENT] key, all the digits flash.
ENT 
020
Touch the [ENT] key again to stop the flashing.
ENT 
C12
Touch the [ENT] key once again to switch to the parameter code selection display.
ENT Basic Touch the [>] key together with the [ENT] key to return to the basic panel display. 
panel (This is not required if you proceed to make another setting.)(The displayed numeric 
display characters indicate the measurement gas concentration.)
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
..
..
.
 Switch operation Display  Description
The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flash in