User Manual

IM 11M13A01-04E
7-6
7.4.3 Entering Parameter Code Selection Display
This section briefly describes the password entry procedure for entering the parameter
code selection display. The password is 1102 - it cannot be changed to a different
password.
Table7.3 Parameter Code Selection
Switch operation Display Description
. ENT
. ENT
. ENT
. ENT
. ENT
. ENT
. ENT
. ENT
. ENT
. ENT
T7.4.3E.EPS
The symbol [ ] indicates that the key is being touched.
Light characters indicates that the digits are flashing.
..
.
.
.
.
.
.
.
.
.
.
.
21.0%
PASSno
0000
1000
1000
1100
1100
1102
1102
A01
Warm-up is complete, and the basic panel is now displayed.
Continuously touch the [ENT] key for at least three seconds
to display "PASSno."
Touch the [ENT] key again. This allows you to change the leftmost
digit that is flashing.
Set the password 1102. If you touch the [ ] key, the digit that is flashing
will be 1.
Touch the [>] key to move the position of the digit that is flashing to the
right one digit.
Touch the [ ] key to change the numeric value to 1.
Touch the [>] key again to move the position of the digit that is flashing
to the right one more digit. Continuously touch the [>] key, and the position
of the digit that is flashing will move continuously to the right.
Touch the [ ] key to change the numeric value to 2. Continuously touch
[>] key, and the numeric value increases continuously
If you touch the [ENT] key, all the digits flash.
Touch the [ENT] key again to display A01 on the parameter-code selection
display.
CAUTION
If no key is touched for at least 20 seconds during password entry, the current display
will automatically switch to the basic panel display.
If no key is touched for at least 10 minutes during parameter code selection, the
current display will automatically switch to the basic panel display.