User Manual

IM 11M13A01-04E
4-2
4.1 Piping for System 1
Piping in System 1 is illustrated in Figure 4.7.
~
ZR202S Integrated type Explosionproof
Zirconia Oxygen Analyzer
F06.EPS
ZA8F flow setting unit
Reference gas
Calibration gas
Needle
valve
Flowmeter
Instrument air
Air Set
Pressure regulator
Zero gas
cylinder
Stop valve
or
Check valve
Span gas (*4)
Non-hazardous Area
Hazardous Area
100 to 240 V AC
Figure 4.7 Piping for System 1
System 1 illustrated in Figure 4.7 requires piping as follows:
• Mount the check valve or the stop valve through a nipple to the calibration gas inlet of
the equipment.
4.1.1 Piping Parts for System 1
Check that the parts listed in Table 4.1 are provided.
Table 4.1 Piping Parts
Equipment Piping location Parts Description
General-use Calibration gas inlet Stop valve or check valve Stop valve (L9852CB or G7016XH
Oxygen Analyzer recommended by YOKOGAWA
Check valve (K9292DN or K9292DS)
provided by YOKOGAWA
Nipple * Rc1/4 or 1/4 NPT Commercially available
Zero gas cylinder User9s scope
Regulator valve (G7013XF or G7014XF) recommended by
YOKOGAWA
Joint for tube connection Rc1/4 or 1/4 NPT Commercially available
Reference gas inlet Air set (K9473XH / K9473XJ or G7004XF
/ K9473XG) recommended by YOKOGAWA
Joint for tube connection Rc1/4 or 1/4 NPT Commercially available
Note: Parts marked with * are used when required.
T4.2E.EPS