Inspection-related Items in Group G
Code Item   Tuning Engineering Default 
 unit   setting
G01 mA-output loop   4 to 20  mA   4 mA
G11 Output contact 1   0: Open  Open
 1: Closed 
G12 Output contact 2   0: Open  Open
 1: Closed 
G15 Automatic calibration 0: Off  Off
 solenoid valve (zero) 1: On  
G16 Automatic calibration 0: Off  Off
 solenoid valve (span) 1: On 
G21 Input1 contact   0: Open
 1: Closed 
G22 Input2 contact   0: Open
 1: Closed 
G30 Reset 
T10.6G.EPS
10.5 Handling of the ZO21S Standard Gas Unit
WARNING
Use only non-hazardous area.
The following describe how to flow zero and span gases using the ZO21S Standard Gas
Unit. Operate the ZO21S Standard Gas Unit, for calibrating a system classified as
System 1, according to the procedures that follow.