10. Other Functions
Equipment Setup and Others in Group F
T10.6F.EPS
Code Item   Tuning   Engineering Default setting 
 unit 
F01 Equipment setup   0: Oxygen analyzer  Oxygen analyzer
 1: Humidity analyzer
F02 Selection of measurement gas   0: Wet  Wet
 1: Dry   
F04 Selection of temperature units   0: degree C
 1: degree F  degree C
F05 Selection of pressure units   0: kPa
 1: psi  kPa
F08 Selection of display items  0: Oxygen concentration  Oxygen concentration
 1: Amount of moisture 
 quantity
 2: Mixed ratio
 3: Item selected with 
 analog output
F10 Date   YY.MM.DD/H
 H.MM 
F11 Period over which average  1 to 255 hours   Hours   One hour
 values are calculated
F12 Period over which max. and   1 to 255 hours   Hours   24 hours
 min. values are monitored
F13 
F14 
F20 Content of moisture in exhaust  0 to 5   m
3
/kg (m
3
) 1.0 m
3
/kg (m
3
) 
 gas 
F21 Theoretical air quantity   0 to 20   m
3
/kg (m
3
) 1.0 m
3
/kg (m
3
)
F22 X value   0 to 19.99   1.0
F23 Ambient absolute  0 to 1   kg/kg   0.1 kg/kg
 humidity 
F30 Initializing all data 
F31 Initializing data in group A 
F32 Initializing data in group B 
F33 Initializing data in group C 
F34 Initializing data in group D 
F35 Initializing data in group E 
F36 Initializing data in group F 