User Manual

IM 11M13A01-04E
10-17
10. Other Functions
Equipment Setup and Others in Group F
T10.6F.EPS
Code Item Tuning Engineering Default setting
unit
F01 Equipment setup 0: Oxygen analyzer Oxygen analyzer
1: Humidity analyzer
F02 Selection of measurement gas 0: Wet Wet
1: Dry
F04 Selection of temperature units 0: degree C
1: degree F degree C
F05 Selection of pressure units 0: kPa
1: psi kPa
F08 Selection of display items 0: Oxygen concentration Oxygen concentration
1: Amount of moisture
quantity
2: Mixed ratio
3: Item selected with
analog output
F10 Date YY.MM.DD/H
H.MM
F11 Period over which average 1 to 255 hours Hours One hour
values are calculated
F12 Period over which max. and 1 to 255 hours Hours 24 hours
min. values are monitored
F13
F14
F20 Content of moisture in exhaust 0 to 5 m
3
/kg (m
3
) 1.0 m
3
/kg (m
3
)
gas
F21 Theoretical air quantity 0 to 20 m
3
/kg (m
3
) 1.0 m
3
/kg (m
3
)
F22 X value 0 to 19.99 1.0
F23 Ambient absolute 0 to 1 kg/kg 0.1 kg/kg
humidity
F30 Initializing all data
F31 Initializing data in group A
F32 Initializing data in group B
F33 Initializing data in group C
F34 Initializing data in group D
F35 Initializing data in group E
F36 Initializing data in group F