User Manual
IM 11M13A01-04E
10-16
Contact-related Items in Group E
Code Item Tuning Engineering Default setting
unit
E01 Selection of input contact 1 0: Invalid Invalid
1: Calibration gas pressure drop
2: Measurement range change
3: Calibration start
4: Detection of non-combusted gas
E02 Selection of input contact 2 0: Invalid Invalid
1: Calibration gas pressure drop
2: Measurement range change
3: Calibration start
4: Detection of non-combusted gas
E03 Selecting action of input contact 1 0: Action with closed contact Action with closed contact
1: Action with open contact
E04 Selecting action of input contact 2 0: Action with closed contact Action with closed contact
1: Action with open contact
E10 Selecting action of output contact 1 0: Action with closed contact (normally Action with closed contact
energized)
1: Action with open contact (normally
deenergized)
E20 Output contact 1 error 0: No action No action
1: Action
E21 Output contact 1, high-high alarm 0: No action No action
1: Action
E22 Output contact 1, high-limit alarm 0: No action No action
1: Action
E23 Output contact 1, low-limit alarm 0: No action No action
1: Action
E24 Output contact 1, low-low alarm 0: No action No action
1: Action
E25 Output contact 1, during maintenance 0: No action Action
1: Action
E26 Output contact 1, during calibration 0: No action No action
1: Action
E27 Output contact 1, measurement range 0: No action No action
change 1: Action
E28 Output contact 1, during warm up 0: No action Action
1: Action
E29 Output contact 1, calibration gas 0: No action No action
pressure decrease 1: Action
E32 Output contact 1, detection of non- 0: No action No action
combusted gas 1: Action
T10.6E.EPS