User Manual

IM 11M13A01-04E
10-14
Output-related Items in Group C
Code Item Tuning Engineering Default setting
unit
C01 Analog output 0: Oxygen concentration Oxygen concentration
1: Amount of moisture content
2: Mixed ratio
C03 Output mode 0: Linear Linear
1: Logarithm
C04 Output during warm-up 0: Held at 4 mA
up 1: Held at 20 mA
2: Set value remains held. Held at 4 mA.
C05 Output during 0: Not held
maintenance 1: Held output just before Held output just before
maintenance service. maintenance service.
2: Set value remains held.
C06 Output during 0: Not held Held output just before
calibration 1: Held output just before calibration.
calibration.
2: Set value remains held.
C07 Output in abnormal 0: Not held Held output at a preset
state 1: Held output just before value.
abnormal state occurs.
2: Set value remains held.
C11 Min. oxygen See Section 8.2. % O
2
0% O
2
concentration
C12 Max. oxygen See Section 8.2. % O
2
25% O
2
concentration
C13
C14
C15
C16
C30 Output damping 0 to 255 Seconds 0 second
factor
C31 Set value during 2.4 to 21.6 mA 4 mA
warm-up
C32 Set value during 2.4 to 21.6 mA 4 mA
maintenance
C33 Set value during 2.4 to 21.6 mA 4 mA
calibration
C34 Set value in abnormal 2.4 to 21.6 mA 3.4 mA
state
T10.6C.EPS