User Manual
IM 11M13A01-04E
10-14
Output-related Items in Group C
Code  Item  Tuning  Engineering Default setting
 unit
C01 Analog output  0: Oxygen concentration  Oxygen concentration
 1: Amount of moisture content
 2: Mixed ratio
C03 Output mode  0: Linear  Linear
 1: Logarithm
C04 Output during warm-up 0: Held at 4 mA
up  1: Held at 20 mA
 2: Set value remains held.  Held at 4 mA.
C05 Output during   0: Not held
maintenance  1: Held output just before  Held output just before
 maintenance service.  maintenance service.
 2: Set value remains held.
C06 Output during   0: Not held  Held output just before
calibration 1: Held output just before  calibration.
 calibration.
 2: Set value remains held.  
C07 Output in abnormal   0: Not held  Held output at a preset
state  1: Held output just before  value.
 abnormal state occurs.
 2: Set value remains held.  
C11 Min. oxygen   See Section 8.2. % O
2 
0% O
2
concentration
C12 Max. oxygen   See Section 8.2. % O
2 
25% O
2
concentration
C13
C14
C15
C16
C30 Output damping   0 to 255  Seconds  0 second
factor
C31 Set value during   2.4 to 21.6 mA  4 mA
warm-up
C32 Set value during   2.4 to 21.6 mA  4 mA
maintenance
C33 Set value during   2.4 to 21.6 mA  4 mA
calibration
C34 Set value in abnormal 2.4 to 21.6 mA  3.4 mA
state
T10.6C.EPS