User Manual
IM 11M13A01-04E
10-13
10. Other Functions
Calibration-related Items in Group B
Code Item Tuning  Engineering unit Default setting
B01 Zero-gas concentration 0.3 to 100  % O
2
 1% O
2
B02 Span-gas concentration 4.5 to 100  % O
2
 21% O
2
B03 Calibration mode 0: Manual calibration    Manual calibration
1: Semi-automatic and manual 
 calibration
2: Automatic, semi-automatic, and 
 manual calibration
B04 Output stabilization time 0 minutes, 0 seconds to 60 minutes, 59   MM.SS  10 minutes, 0 seconds
seconds
B05 Calibration time 0 minutes, 0 seconds to 60 minutes, 59  MM.SS  10 minutes, 0 seconds
seconds
B06 Calibration period 0 days 0 hours to 255 days 23 hours  DD.HH  30 days, 0 hours
B07 Start time  YY.MM.DD/HH.MM 00.01.01.00.00
B08 Calibration  0: Zero and span  Zero and span
1: Span only
2: Zero only
B09 Calibration concentration  Display only  % O
2
measurement
B10 Manual calibration
B11 Semi-automatic calibration
T10.6B.EPS










