User Manual
IM 11M13A01-04E
9-9
9. Calibration
9.2.2 Calibration
9.2.2.1Manual Calibration
For manual calibration, consult Section 7.10, “Calibration” earlier in this manual.
9.2.2.2Semi-automatic Calibration
(1) Calibration startup using infrared switches
Table 9.3 Semi-automatic Calibration Procedure
 Switch operation   Display  Brief description 
>
ENT 
b11
>
ENT 
SA-CAL
>
ENT 
SPAn
 /20.84
>
ENT 
ZEro
/0.89
>
ENT 
CALEnd
>
ENT 
T9.3E.EPS
Change the parameter code to B11. (Previous operations omitted)
Touch the [ENT] key to display "SA-CAL" (Semi Auto CAL).
Touch the [ENT] key again to open the span-gas solenoid valve. 
The span gas then flows. "SPAN" and the currently measured value are alternately 
displayed. If the "output hold" is set, the output hold will start at this time.
If the set calibration time elapses, the span-gas solenoid valve closes automatically, 
the zero-gas solenoid valve opens and the zero gas flows. "ZERO" and the currently 
measured value are displayed alternately.
End If the set calibration time elapses, the zero-gas solenoid valve then closes 
automatically. The ’’CALEND’’ flashes until the set output stabilization time elapses.
If the output stabilization time elapses, the basic panel display then appears. 
Output holding will be released.
.
.....
Basic panel 
display
The symbol ( ) indicates that the key is being touched. Light characters indicate that the digits are flashing.
/ indicates that both are displayed alternately.
(2) To start calibration using an input contact, follow these steps:
• Make sure that Calibration start has been selected in the Input contacts display (see
Section 8.4, earlier in this manual).
• Apply an input contact to start calibration.
(3) To stop calibration midway, follow these steps:
(1) Touch the [>] key and [ENT] key together. The calibration will stop and the
output stabilization time will be set up.
(4) Touch the [>] key once again to return to the basic panel display and the analyzer
will be in normal measurement.
9.2.2.3Automatic Calibration
No execution operations are required for automatic calibration. Automatic calibration
starts in accordance with a preset start date and time. Calibration is then executed at
preset intervals.
CAUTION
Before conducting a semi-automatic or automatic calibration, run the automatic calibra-
tion unit beforehand to obtain a calibration flow of 600 6 60 ml/min.










