Owner's manual

<8. Detailed Data Setting>
8-1
IM 11M12A01-04E
8. Detailed Data Setting
8.1 Setting Display Item
Display items are dened as items displayed on the basic panel display.
Parameter code A00 or F08 is used to set the display items as shown in Table 8.1. The oxygen
concentration is set at the factory before shipment. In addition, if the data initialization is performed,
the oxygen concentration will be set.
Table 8.1 Display Item
Values set with A00 or F08 Items displayed on the basic panel display
0 Indicates the oxygen concentration.
1 or 2 For humidity analyzers only. (if 1 or 2 is set for the oxygen analyzer,
"0.0" is only displayed on the basic panel display.)
3 Displays an item for the current output.
If the output damping has been set for the current output, values
involving the output damping are displayed.
CAUTION
If you set “3in the parameter code A00 or F08, be sure to select “Oxygen Concentrationin the
following mA output setting (see Section 8.2,Current Output Setting”).
8.2 Current Output Setting
This section describes setting of the analog output range. Table 8.2 shows parameter codes for the
set items.
Table 8.2 Current Output Parameter Codes
Set item Parameter code Set value
Analog output C01 0 Oxygen concentration
1 4 mA (xed *1)
2 4 mA (xed *1)
Output mode C03 0 Linear
1 Logarithm
Min. oxygen concentration C11 Oxygen concentration at 4 mA
Max. oxygen concentration C12 Oxygen concentration at 20 mA
Output damping constant C30 0 to 255 seconds
*1: For the oxygen analyzer, set 0 (zero) only for parameter code C01. When it is set, the current output is 4-mA xed regardless of
the oxygen concentration.