Owner's manual

<7. Startup>
7-16
IM 11M12A01-04E
Table 7.12 Calibration Setup Procedure
Switch operation Display Description
>
ENT
A01
Display after the password has been entered.
>
ENT
b01
Set the zero gas concentration. Switch the parameter code to B01.
Here, set 0.98%.
>
ENT
001.00
%
Touch the [ENT] key to display the currently set value.
>
ENT
001.00
%
Touch the [>] key to move the position of the digit that is ashing to 1.
>
ENT
000.00
%
Touch the [] key to change to 0.
>
ENT
000.00
%
Touch the [>] key to move the position of the digit that is ashing to the
right one digit.
>
ENT
000.90
%
Touch the [] key to change the numeric value to 9.
>
ENT
000.90
%
Touch the [>] key to move the position of the digit that is ashing to the
right one digit.
>
ENT
000.98
%
Touch the [] key to change the numeric value to 8.
>
ENT
000.98
%
Touch the [ENT] key to have all the digits ash.
>
ENT
000.98
%
Touch the [ENT] key again to stop the ashing.
>
ENT
b01
Touch the [ENT] key once again to switch to the parameter code selection
display.
Set the span gas concentration by above procedure, set 21 %.
>
ENT
b03
Next, set the calibration mode. Switch the parameter code to B03.
>
ENT
0
Touch the [ENT] key to display the currently set value.
If it is 0, you can leave it as is. If it is other than 0, change it to 0 (zero).
>
ENT
0
Touch the [ENT] key. The numeric value will ash.
>
ENT
0
Touch the [ENT] key again to stop the ashing.
>
ENT
b03
Touch the [ENT] key once again to switch to the parameter code selection
display.
>
ENT Basic panel
display
Touch the [>] key together with the [ENT] key to return to the basic panel
display. (This is not required if you proceed to make another setting.)
(The displayed numeric characters indicate the measurement gas
concentration.)
The symbol [ ] indicates that the key is being touched. Light characters indicates that the digits are ashing.